Literature DB >> 10973411

Missing anterior cervical plate and screws: a case report.

S Fujibayashi1, J Shikata, N Kamiya, C Tanaka.   

Abstract

STUDY
DESIGN: A case report of an anterior cervical plate and screws that disappeared completely.
OBJECTIVES: To present a case of a missing anterior cervical plate and screws, this being quite a rare complication of a cervical implant. SUMMARY OF BACKGROUND DATA: No mention of this complication was found in the literature.
METHODS: Methods in the literature and clinical presentation are reviewed.
RESULTS: Dislodgment of an anterior cervical plate occurred in association with an methicillin-resistant Staphylococcus aureus infection. Three months later the implants had disappeared, presumably passing without notice through the gastrointestinal tract. Whole body fluoroscopy could not identify any residual plate or screws. No esophageal fistula could be detected by barium meal swallow study.
CONCLUSIONS: The plate and screws became dislodged, eroded through the posterior wall of the esophagus, then traversed the limit of the gastrointestinal tract, passing with feces without significant morbidity. Whether the methicillin-resistant Staphylococcus aureus infection contributed to this unique circumstance or is coincidentally related to it remained a matter of speculation.

Entities:  

Mesh:

Year:  2000        PMID: 10973411     DOI: 10.1097/00007632-200009010-00018

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  16 in total

1.  Late hypopharyngo-esophageal perforation after cervical spine surgery: proposal of a therapeutic strategy.

Authors:  Burkhard H A von Rahden; Hubert J Stein; Michael A Scherer
Journal:  Eur Spine J       Date:  2005-09-07       Impact factor: 3.134

2.  Erosion of anterior cervical plate into pharynx with pharyngotracheal fistula.

Authors:  Yuo-Chen Kuo; Marc S Levine
Journal:  Dysphagia       Date:  2010-01-21       Impact factor: 3.438

3.  A comparison of a new zero-profile, stand-alone Fidji cervical cage and anterior cervical plate for single and multilevel ACDF: a minimum 2-year follow-up study.

Authors:  Zhonghai Li; Yantao Zhao; Jiaguang Tang; Dongfeng Ren; Jidong Guo; Huadong Wang; Li Li; Shuxun Hou
Journal:  Eur Spine J       Date:  2016-08-23       Impact factor: 3.134

4.  The application of zero-profile anchored spacer in anterior cervical discectomy and fusion.

Authors:  Zhiwen Wang; Weimin Jiang; Xuefeng Li; Heng Wang; Jinhui Shi; Jie Chen; Bin Meng; Huilin Yang
Journal:  Eur Spine J       Date:  2014-10-22       Impact factor: 3.134

5.  Expectoration of anterior cervical discectomy and fusion cage: a case report.

Authors:  Parker J Prusick; Shahbaaz A Sabri; Christopher J Kleck
Journal:  J Spine Surg       Date:  2021-06

6.  Delayed anterior cervical plate dislodgement with pharyngeal wall perforation and oral extrusion of cervical plate screw after 8 years: A very rare complication.

Authors:  Ravindranath Kapu; Manish Singh; Anil Pande; Matabushi Chakravarthy Vasudevan; Ravi Ramamurthi
Journal:  J Craniovertebr Junction Spine       Date:  2012-01

7.  Delayed esophageal pseudodiverticulum after anterior cervical spine fixation: report of 2 cases.

Authors:  Ali Sadrizadeh; Ehsan Soltani; Mehdi Abili; Paria Dehghanian
Journal:  Iran J Otorhinolaryngol       Date:  2015-03

8.  Missing screw as a rare complication of anterior cervical instrumentation.

Authors:  Yusuf Kurtuluş Duransoy; Mesut Mete; Baha Zengel; Mehmet Selçukı
Journal:  Case Rep Orthop       Date:  2013-03-24

9.  A rare case of hypopharyngeal screw migration after spine stabilization with plating.

Authors:  G Salis; B Pittore; G Balata; C Bozzo
Journal:  Case Rep Otolaryngol       Date:  2013-06-09

10.  Integrated intervertebral device for anterior cervical fusion: An initial experience.

Authors:  Manish K Kasliwal; John E O'toole
Journal:  J Craniovertebr Junction Spine       Date:  2012-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.