| Literature DB >> 10972155 |
Abstract
The X-ray emission probabilities of the Mn-K(alpha) and Mn-K(beta) radiation in the decay of 55Fe were measured with a calibrated Si(Li) spectrometer, resulting in 0.245(7) and 0.0338(9), respectively. A half-life of 1003.5(21) day was measured using the same detector.Mesh:
Substances:
Year: 2000 PMID: 10972155 DOI: 10.1016/s0969-8043(00)00166-4
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513