Alterations in coagulation after implantation of a pulsatile Novacor LVAD and the axial flow MicroMed DeBakey LVAD.

A Koster1, M Loebe, R Hansen, E V Potapov, G P Noon, H Kuppe, R Hetzer.   

Abstract

BACKGROUND: The MicroMed DeBakey left ventricular assist device (LVAD) is a chamber and valveless axial flow blood pump. We investigated parameters of the coagulation system in patients after implantation of the axial flow LVAD and patients following implantation of a pulsatile Novacor LVAD.
METHODS: Six consecutive patients of both groups were investigated over a period of 6 weeks after implantation. beta-Thromboglobulin, platelet factor 4, factor XIIa, thrombin/antithrombin complexes, plasmin/alpha2-antiplasmin complexes, and D-Dimer levels were measured.
RESULTS: With the exception of the plasmin/alpha2-antiplasmin levels in the Novacor group, all parameters were elevated in both groups. The levels of beta-thromboglobulin, platelet factor 4, factor XIIa, and plasmin/alpha2-antiplasmin were significantly increased in the axial flow LVAD group.
CONCLUSIONS: The axial flow LVAD strongly influences the systems of contact activation and fibrinolysis. The elevation of platelet proteins appears to follow platelet damage. Although no thromboembolic events were observed in both groups, elevation of thrombin/antithrombin complexes provides convincing evidence of an increased activation of the coagulation system and the concomitant risk for the development of thromboembolism.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10969675     DOI: 10.1016/s0003-4975(00)01404-1

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  6 in total

Review 1.  The biological basis of thrombosis and bleeding in patients with ventricular assist devices.

Authors:  Ranjit John; Sangjin Lee
Journal:  J Cardiovasc Transl Res       Date:  2008-10-17       Impact factor: 4.132

Review 2.  Plasma contact factors as therapeutic targets.

Authors:  Benjamin F Tillman; Andras Gruber; Owen J T McCarty; David Gailani
Journal:  Blood Rev       Date:  2018-04-12       Impact factor: 8.250

Review 3.  Inhibition of Factors XI and XII for Prevention of Thrombosis Induced by Artificial Surfaces.

Authors:  Benjamin Tillman; David Gailani
Journal:  Semin Thromb Hemost       Date:  2017-09-12       Impact factor: 4.180

Review 4.  Mechanical blood trauma in assisted circulation: sublethal RBC damage preceding hemolysis.

Authors:  Salim E Olia; Timothy M Maul; James F Antaki; Marina V Kameneva
Journal:  Int J Artif Organs       Date:  2016-03-30       Impact factor: 1.595

5.  Moderation of prekallkrein-factor XII interactions in surface activation of coagulation by protein-adsorption competition.

Authors:  Kaushik Chatterjee; Jennifer L Thornton; James W Bauer; Erwin A Vogler; Christopher A Siedlecki
Journal:  Biomaterials       Date:  2009-06-23       Impact factor: 12.479

Review 6.  Bleeding with the artificial heart: Gastrointestinal hemorrhage in CF-LVAD patients.

Authors:  Grigoriy E Gurvits; Elena Fradkov
Journal:  World J Gastroenterol       Date:  2017-06-14       Impact factor: 5.742