| Literature DB >> 10966182 |
N Hata1, A Nabavi, W M Wells, S K Warfield, R Kikinis, P M Black, F A Jolesz.
Abstract
A three-dimensional optical flow method to measure volumetric brain deformation from sequential intraoperative MR images and preliminary clinical results from five cases are reported. Intraoperative MR images were scanned before and after dura opening, twice during tumor resection, and immediately after dura closure. The maximum cortical surface shift measured was 11 mm and subsurface shift was 4 mm. The computed deformation field was most satisfactory when the skin was segmented and removed from the images before the optical flow computation.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10966182 DOI: 10.1097/00004728-200007000-00004
Source DB: PubMed Journal: J Comput Assist Tomogr ISSN: 0363-8715 Impact factor: 1.826