E A Shirtcliff1, D A Granger, E B Schwartz, M J Curran, A Booth, W H Overman.
Abstract
We developed simple, reliable, and highly sensitive assay modifications of commercially available radioimmunoassay kits to measure estradiol in saliva and blood spot specimens. The saliva assay has average intra- and interassay coefficients of variation (CV) of 6.45 and 9.01%, with average analytical and serial dilution recoveries 100.65 and 89.25%. The blood spot assay has average intra- and interassay CVs of 7.57 and 8.22%, with analytical and serial dilution recoveries of 80.50 and 108.50%. The analytical sensitivity ranges of the saliva (0.25-7.50 pg/ml) and blood spot (2. 00-375 pg/ml) assays are sufficient to determine levels in the majority of pre- and postpubertal males and females. Blood spot assay results are correlated with serum estradiol levels for adult males, r (17) = 0.73, and females, r (18) = 0.96. In contrast, the serum-saliva correlation is only modest for adult females, r (14) = 0.60, and not significant for adult males. Substitution of blood spot assay results for serum values underestimates the known serum estradiol-behavior correlation by only 3.45%, whereas substitution of saliva assay results for serum values underestimates the association by 37.55%. The findings have important implications for the use and potential misuse of noninvasive measures of estradiol in studies of health and human development. Copyright 2000 Academic Press.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10964528 DOI: 10.1006/hbeh.2000.1614
Source DB: PubMed Journal: Horm Behav ISSN: 0018-506X Impact factor: 3.587