The strength-duration curve and its importance in pacing efficiency: a study of 325 pacing leads in 229 patients.

S Coates1, B Thwaites.   

Abstract

Pacemaker battery life is dependent on programmable parameters, principally pulse amplitude and pulse duration. High factory default settings cause excessive current drain. The strength-duration curve relates pacing threshold to pulse duration. The most energy efficient pacing occurs at chronaxie, a value of pulse duration derived from the curve. Strength-duration curves were calculated for 325 acutely implanted pacing leads. Chronaxie and rheobase were compared for atrial and ventricular leads. Chronaxie was compared with actual programmed pulse duration. There were 101 atrial and 224 ventricular leads, all passive fixation. The curve fit was good, (mean error +/- SD) 0.024 +/- 0.06 V for atrial curves and 0.008 +/- 0.034 V for ventricular curves. Mean (+/- SD) atrial and ventricular chronaxies were 0.24 +/- 0.07 ms and 0.25 +/- 0.07 ms, respectively. A "Z" value of 1.4 indicated that chronaxies might have been from the same population. Mean (+/- SD) atrial and ventricular rheobases were 0.51 +/- 0.2 V and 0.35 +/- 0.13 V, respectively. A "Z" value of 7.1 (P < 0.001) suggested atrial and ventricular rheobases were from differing populations. All patients had factory default pulse durations of 0.45 ms or 0.5 ms, exceeding acute chronaxie by a factor of two, thus, demonstrating suboptimal pacing. We conclude that understanding the strength-duration curve is critical. Sensible programming of other pacing functions optimizes longevity. Battery drain is reduced by programming pulse duration to chronaxie with a doubling of voltage threshold at this point to achieve a safety margin. Further study of chronaxie drift with time is required.

Entities:  

Mesh:

Year:  2000        PMID: 10962751     DOI: 10.1111/j.1540-8159.2000.tb00943.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Intraoperative sensing increase predicts long-term pacing threshold in leadless pacemakers.

Authors:  Gianfranco Mitacchione; Gianmarco Arabia; Marco Schiavone; Manuel Cerini; Alessio Gasperetti; Francesca Salghetti; Luca Bontempi; Maurizio Viecca; Antonio Curnis; Giovanni B Forleo
Journal:  J Interv Card Electrophysiol       Date:  2022-01-04       Impact factor: 1.900

2.  Disparity in left ventricular stimulation among different pacing configurations in cardiac resynchronization therapy.

Authors:  Jose F Huizar; Karoly Kaszala; Jayanthi Koneru; Marcin Kowalski; Leroy R Thacker; Mark A Wood; Kenneth A Ellenbogen
Journal:  Circ Arrhythm Electrophysiol       Date:  2011-12-22

3.  Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study.

Authors:  Philippe Ritter; Gabor Z Duray; Clemens Steinwender; Kyoko Soejima; Razali Omar; Lluís Mont; Lucas V A Boersma; Reinoud E Knops; Larry Chinitz; Shu Zhang; Calambur Narasimhan; John Hummel; Michael Lloyd; Timothy Alexander Simmers; Andrew Voigt; Verla Laager; Kurt Stromberg; Matthew D Bonner; Todd J Sheldon; Dwight Reynolds
Journal:  Eur Heart J       Date:  2015-06-04       Impact factor: 29.983

4.  Energy-efficiency of Cardiomyocyte Stimulation with Rectangular Pulses.

Authors:  Martin Laasmaa; Pengfei Lu; Mladen Veletić; William E Louch; Jacob Bergsland; Ilangko Balasingham; Marko Vendelin
Journal:  Sci Rep       Date:  2019-09-16       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.