Literature DB >> 10960990

The effect of a hydroxyapatite-reinforced polyethylene stress distributor in a dental implant on compressive stress levels in surrounding bone.

O A Abu-Hammad1, A Harrison, D Williams.   

Abstract

This study investigated various designs of stress breakers incorporated into the dental implant using 3-D finite element analysis. These designs employed hydroxyapatite-reinforced polyethylene (HRP), a material capable of inducing osseointegration. The most successful design was that of a dental implant with a peripheral HRP component that was in direct contact with the bone surrounding the neck of the implant. This design lowered the compressive stress values in bone around the neck of the implant. Attempts were also made to optimize this design.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10960990

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  2 in total

1.  A Finite Element Stress Analysis of a Concical Triangular Connection in Implants: A New Proposal.

Authors:  Romy Angeles Maslucan; John Alexis Dominguez
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

2.  Numerical study of effect of elastomeric stress absorbers on stress reduction in bone-dental implant interface.

Authors:  Ghalem Mehdi; Abderrahmane Belarbi; Bensmaine Mansouri; Zitouni Azari
Journal:  J Appl Oral Sci       Date:  2015 Jan-Feb       Impact factor: 2.698

  2 in total