Constrained treatment planning using sequential beam selection.

E Woudstra1, P R Storchi.   

Abstract

In this paper an algorithm is described for automated treatment plan generation. The algorithm aims at delivery of the prescribed dose to the target volume without violation of constraints for target, organs at risk and the surrounding normal tissue. Pre-calculated dose distributions for all candidate orientations are used as input. Treatment beams are selected in a sequential way. A score function designed for beam selection is used for the simultaneous selection of beam orientations and weights. In order to determine the optimum choice for the orientation and the corresponding weight of each new beam, the score function is first redefined to account for the dose distribution of the previously selected beams. Addition of more beams to the plan is stopped when the target dose is reached or when no additional dose can be delivered without violating a constraint. In the latter case the score function is modified by importance factor changes to enforce better sparing of the organ with the limiting constraint and the algorithm is run again.

Mesh:

Year:  2000        PMID: 10958185     DOI: 10.1088/0031-9155/45/8/306

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

Review 1.  Recent developments in non-coplanar radiotherapy.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; James L Bedford