Abstract

STATEMENT OF PROBLEM: Implant abutment screw joints tend to loosen under clinical conditions. During impression and prosthesis fabrication, repeated clinical closing and opening of abutment screws may cause component wear and decrease frictional fit of the mating parts, resulting in altered resistance to opening and potential for loss of preload in function.
PURPOSE: This study recorded changes in opening torque values due to multiple consecutive closures at a constant torque within and between different abutment/implant (A/I) systems.
MATERIAL AND METHODS: Repeated opening-closing cycles were used to simulate in vitro embedment relaxation and component wear of 7 A/I systems from 5 manufacturers. Screw opening torque values were recorded up to 200 consecutive closures at 20 N/cm.
RESULTS: A progressive decrease in opening torque values was measured in all implant systems. Significant differences were found between A/I systems. Systems with morse tapered and spline connections consistently maintained a higher resistance to opening force. Percentage torque loss ranged from 3% to 20% on immediate opening, and from 4.5% to 36% for average of first 30 opening/closing cycles.
CONCLUSION: Repeated opening and closing of implant abutment screws caused progressive loss of torque retention with variations between systems. This was probably due to a decrease in the coefficient of friction between the mating components.

Mesh:

Substances:

Year:  2000        PMID: 10946337     DOI: 10.1067/mpr.2000.108069

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  15 in total

1.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

2.  Loosening torque of Universal Abutment screws after cyclic loading: influence of tightening technique and screw coating.

Authors:  Atais Bacchi; Alexandre Regalin; Claudia Lopes Brilhante Bhering; Rodrigo Alessandretti; Aloisio Oro Spazzin
Journal:  J Adv Prosthodont       Date:  2015-10-28       Impact factor: 1.904

3.  Abrasion of abutment screw coated with TiN.

Authors:  Seok-Won Jung; Mee-Kyoung Son; Chae-Heon Chung; Hee-Jung Kim
Journal:  J Adv Prosthodont       Date:  2009-07-31       Impact factor: 1.904

4.  Effect of cyclic loading and retightening on reverse torque value in external and internal implants.

Authors:  Woong-Rae Cho; Yoon-Hyuk Huh; Chan-Jin Park; Lee-Ra Cho
Journal:  J Adv Prosthodont       Date:  2015-08-18       Impact factor: 1.904

5.  Stability of external and internal implant connections after a fatigue test.

Authors:  Paulo Cesar Pinheiro Feitosa; Ana Paula Barbosa de Lima; Laís Regiane Silva-Concílio; William Cunha Brandt; Ana Christina Claro Neves
Journal:  Eur J Dent       Date:  2013-07

6.  A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants.

Authors:  Felipe Miguel Saliba; Mayra Cardoso; Marcelo Ferreira Torres; Alexandre Carvalho Teixeira; Eduardo José Veras Lourenço; Daniel de Moraes Telles
Journal:  J Appl Oral Sci       Date:  2011 Jan-Feb       Impact factor: 2.698

7.  Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants.

Authors:  Hnd Hadi Mohammed; Jin-Han Lee; Ji-Myung Bae; Hye-Won Cho
Journal:  J Adv Prosthodont       Date:  2016-02-23       Impact factor: 1.904

8.  Stability of the prosthetic screws of three types of craniofacial prostheses retention systems.

Authors:  Antonio Gabriel Lanata-Flores; Eder Alberto Sigua-Rodriguez; Douglas Rangel Goulart; Veber Luiz Bomfim-Azevedo; Sergio Olate; José Ricardo de Albergaria-Barbosa
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2016-12-27

9.  Microbiological Sealing Analysis of a Tapered Connection and External Hexagon System.

Authors:  Gardel Nepomuceno Costa; Elizabeth Ferreira Martinez; Aluísio Martins de Oliveira Ruellas; Daiane Cristina Peruzzo; Júlio Cesar Joly; Marcelo Henrique Napimoga
Journal:  Int J Dent       Date:  2017-02-28

10.  Influence of the implant abutment types and the dynamic loading on initial screw loosening.

Authors:  Eun-Sook Kim; Soo-Yeon Shin
Journal:  J Adv Prosthodont       Date:  2013-02-28       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.