Literature DB >> 10943757

Portable dynamometer anchoring station for measuring strength of the hip extensors and abductors.

S F Nadler1, M L DePrince, N Hauesien, G A Malanga, T P Stitik, E Price.   

Abstract

OBJECTIVE: To assess the reliability of a specially designed dynamometer anchoring station in the measurement of hip muscle strength.
DESIGN: Prospective study using test-retest design.
SETTING: Outpatient clinic setting. PARTICIPANTS: Ten subjects, ages 25 to 35yrs. MAIN OUTCOME MEASURES: Using the apparatus, three consecutive measures were recorded for hip abduction and extension by an inexperienced examiner. Two weeks later, the same subjects were retested, with the evaluator blinded to the initial results. In both the test and retest, average and maximal values of strength were calculated for each muscle. Reliability of the device was assessed by computing the intraclass correlation coefficients and coefficients of variation (CVs).
RESULTS: Intraclass correlation coefficients ranged from .94 to .98. The average CV for the maximal and average abduction strength had values of 4.77% and 4%, respectively. The maximal and average extension strength had average CV values of 8.06% and 7.83%, respectively.
CONCLUSION: This specially designed dynamometer anchoring station has been found to be highly reliable in the measurement of hip girdle strength and has the advantage of easy adjustability and portability for large-scale screenings. This device enhances the reliability of the dynamometer, which may be subject to considerable variability when applied by hand to the powerful hip girdle musculature.

Entities:  

Mesh:

Year:  2000        PMID: 10943757     DOI: 10.1053/apmr.2000.7165

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  20 in total

1.  Precision, comfort and mechanical performance of the Quadriso-tester, a quadriceps force measuring device.

Authors:  G J Verkerke; K A P M Lemmink; A J Slagers; M H Westhoff; G A J van Riet; G Rakhorst
Journal:  Med Biol Eng Comput       Date:  2003-05       Impact factor: 2.602

2.  Reliability of stabilised commercial dynamometers for measuring hip abduction strength: a pilot study.

Authors:  P Click Fenter; J W Bellew; T A Pitts; R E Kay
Journal:  Br J Sports Med       Date:  2003-08       Impact factor: 13.800

3.  Hip- and knee-strength assessments using a hand-held dynamometer with external belt-fixation are inter-tester reliable.

Authors:  Kristian Thorborg; Thomas Bandholm; Per Hölmich
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-07-07       Impact factor: 4.342

4.  Calculation of Resistive Loads for Elastic Resistive Exercises.

Authors:  Kelsey Picha; Tim Uhl
Journal:  J Sport Rehabil       Date:  2018-03-14       Impact factor: 1.931

5.  The reliability of portable fixed dynamometry during hip and knee strength assessments.

Authors:  Roger O Kollock; James A Onate; Bonnie Van Lunen
Journal:  J Athl Train       Date:  2010 Jul-Aug       Impact factor: 2.860

6.  Elastic Resistance Effectiveness on Increasing Strength of Shoulders and Hips.

Authors:  Kelsey J Picha; Muataz R Almaddah; Jordan Barker; Tavis Ciochetty; W Scott Black; Tim L Uhl
Journal:  J Strength Cond Res       Date:  2019-04       Impact factor: 3.775

7.  Evaluation of peripheral muscle strength of patients undergoing elective cardiac surgery: a longitudinal study.

Authors:  Kelli Maria Souza Santos; Manoel Luiz de Cerqueira Neto; Vitor Oliveira Carvalho; Valter Joviniano de Santana Filho; Walderi Monteiro da Silva Junior; Amaro Afrânio Araújo Filho; Telma Cristina Fontes Cerqueira; Lucas de Assis Pereira Cacau
Journal:  Rev Bras Cir Cardiovasc       Date:  2014 Jul-Sep

8.  Skeletal Muscle and Childhood Cancer: Where are we now and where we go from here.

Authors:  Chelsea G Goodenough; Robyn E Partin; Kirsten K Ness
Journal:  Aging Cancer       Date:  2021-05-20

9.  Correlations among multiple measures of functional and mechanical instability in subjects with chronic ankle instability.

Authors:  Tricia J Hubbard; Lauren C Kramer; Craig R Denegar; Jay Hertel
Journal:  J Athl Train       Date:  2007 Jul-Sep       Impact factor: 2.860

10.  Reliability of a new stabilized dynamometer system for the evaluation of hip strength.

Authors:  Timothy W Tourville; Helen C Smith; Sandra J Shultz; Pamela M Vacek; James R Slauterbeck; Robert J Johnson; Bruce D Beynnon
Journal:  Sports Health       Date:  2013-03       Impact factor: 3.843

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.