K Nithipatikom1, P F Pratt, W B Campbell.
Abstract
Epoxyeicosatrienoic acids (EETs) are cytochrome P-450 metabolites of arachidonic acid involved in the regulation of vascular tone. The method of microbore column high-performance liquid chromatography with fluorescence detection was developed to determine 14,15-EET, 11, 12-EET, and the mixture of 8,9-EET and 5,6-EET. Tridecanoic acid (TA) was used as an internal standard. EETs were reacted with 2-(2, 3-naphthalimino)ethyl trifluoromethanesulfonate (NT) to form highly fluorescent derivatives. A C(18) microbore column and a water-acetonitrile mobile phase were used for separation. Samples were excited at 259 nm, and the fluorescence was detected at 395 nm. The overall recoveries were 88% for EETs and 40% for TA. EETs were detected in concentrations as low as 2 pg (signal-to-noise ratio = 3). The method was used to determine the EET production from endothelial cells (ECs). Bradykinin and methacholine (10(-6) M) stimulated an increase in the production of EETs by ECs two- and fivefold, respectively. This sensitive method may be used for determination of EETs at low concentrations normally detected in complex biological samples.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10924087 DOI: 10.1152/ajpheart.2000.279.2.H857