OBJECTIVE: The purpose of this study was to determine the frequency of apparent stenosis of normal aortic branches in patients on first-pass gadolinium-enhanced three-dimensional (3D) MR angiography and to reproduce the same phenomenon in a pulsatile flow phantom model. CONCLUSION: Apparent stenosis of normal vessels on gadolinium-enhanced 3D MR angiography seen on the first-pass acquisition was observed in only a small proportion (approximately 2%) of our patients. The pseudostenosis was reproducible in the phantom model using rapid injection. A stenosis on first-pass images should be interpreted with caution. Confirmation of the findings on other sequences, such as the second-pass gadolinium-enhanced 3D MR angiography or 3D phase-contrast MR angiography, prevented overdiagnosis of significant stenoses.
OBJECTIVE: The purpose of this study was to determine the frequency of apparent stenosis of normal aortic branches in patients on first-pass gadolinium-enhanced three-dimensional (3D) MR angiography and to reproduce the same phenomenon in a pulsatile flow phantom model. CONCLUSION: Apparent stenosis of normal vessels on gadolinium-enhanced 3D MR angiography seen on the first-pass acquisition was observed in only a small proportion (approximately 2%) of our patients. The pseudostenosis was reproducible in the phantom model using rapid injection. A stenosis on first-pass images should be interpreted with caution. Confirmation of the findings on other sequences, such as the second-pass gadolinium-enhanced 3D MR angiography or 3D phase-contrast MR angiography, prevented overdiagnosis of significant stenoses.
Authors: Achim Seeger; Ulrich Kramer; Michael Fenchel; Florian Grimm; Christiane Bretschneider; Jörg Döring; Bernhard Klumpp; Gunnar Tepe; Kilian Rittig; Peter R Seidensticker; Claus D Claussen; Stephan Miller Journal: J Cardiovasc Magn Reson Date: 2008-12-30 Impact factor: 5.364