Literature DB >> 10912693

Magnetic resonance imaging of stapes prostheses.

A J Syms1, G W Petermann.   

Abstract

OBJECTIVE: To assess the magnetic resonance imaging (MRI) compatibility of stapes prostheses. DATA SOURCES: A comprehensive review of the English literature evaluating MRI compatibility of stapes prostheses. Multiple series of stapes prostheses from different manufactures placed in a 1.5-tesla MRI field to determine ferromagnetic properties.
RESULTS: When MRI was first introduced, reports demonstrated the MRI compatibility of stapes prostheses. The testing was performed on single copies of various prostheses from several manufacturers. Although implant manufacturers have indicated MRI compatibility, several reports of variable ferromagnetic properties of aneurysm clips have been reported. This variability has led to rotation of the clips and hemorrhage in patients with supposed MRI-compatible clips. These findings suggest that testing single stapes prostheses from a manufacturer might not completely assess the safety of MRI on patients with stapes prostheses. We performed MRI compatibility testing on several series of stapes prostheses from Xomed Surgical Products and Smith & Nephew Richards. Two series of Xomed stapes prostheses were found to have ferromagnetic properties.
CONCLUSION: Manufacturing variability could lead to stapes prostheses being MRI incompatible. Each prosthesis should be tested before implantation for ferromagnetic properties.

Entities:  

Mesh:

Year:  2000        PMID: 10912693

Source DB:  PubMed          Journal:  Am J Otol        ISSN: 0192-9763


  2 in total

Review 1.  Imaging of the postoperative middle ear.

Authors:  Marc T Williams; Denis Ayache
Journal:  Eur Radiol       Date:  2004-01-29       Impact factor: 5.315

2.  Behavior of metal implants used in ENT surgery in 7 Tesla magnetic resonance imaging.

Authors:  Ariane Thelen; Hans-Christian Bauknecht; Patrick Asbach; Thomas Schrom
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-07-12       Impact factor: 2.503

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.