Ability of a confocal scanning laser ophthalmoscope (TopSS) to detect early glaucomatous visual field defect.

B S Ahn1, C Kee.   

Abstract

AIMS: To evaluate the ability of the confocal scanning laser ophthalmoscope (TopSS) to detect early glaucomatous visual field defect using our unique discriminant criteria.
METHODS: The optic discs of 110 eyes of normal Korean subjects were examined and normal values for each variable were obtained according to the size of the optic disc. The five most sensitive optic disc variables for discriminating glaucoma were then applied to one eye of 80 Korean subjects with primary open angle glaucoma or normal tension glaucoma. Only eyes with an optic disc size of 2.0-3.0 mm(2) and a contour tilt of less than 3 degrees were included. These variables were used to develop unique discriminant criteria for detecting early glaucomatous visual field defect and their sensitivity and specificity were calculated in three groups of patients with visual field loss.
RESULTS: The five most sensitive variables were half the depth area, cup/disc (C/D) ratio, total area of the neuroretinal rim (NRR), volume above, and localised thinning of the NRR. The following criteria were used to diagnose glaucoma: (1) total area of the NRR decreased and one of the other four variables abnormal, and (2) total area of the NRR normal, localised thinning of the NRR, and one of the other three variables abnormal. The sensitivity of these criteria was 89.7% in patients with a mildly impaired visual field and 100% in those with a moderately or severely impaired visual field; the specificity was 89.1%.
CONCLUSIONS: The discriminant criteria used had high sensitivity and specificity in the diagnosis of glaucoma and the TopSS can be useful in the early detection of changes in the glaucomatous optic disc.

Entities:  

Mesh:

Year:  2000        PMID: 10906090      PMCID: PMC1723587          DOI: 10.1136/bjo.84.8.852

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  26 in total

1.  Effect of optic disc size or age on evaluation of optic disc variables.

Authors:  C Kee; H Koo; Y Ji; S Kim
Journal:  Br J Ophthalmol       Date:  1997-12       Impact factor: 4.638

2.  Reproducibility of topographic measures of the glaucomatous optic nerve head.

Authors:  O Geyer; A Michaeli-Cohen; D M Silver; D Versano; M Neudorfer; R Dzhanov; M Lazar
Journal:  Br J Ophthalmol       Date:  1998-01       Impact factor: 4.638

3.  The effect of optic disc size on diagnostic precision with the Heidelberg retina tomograph.

Authors:  M Iester; F S Mikelberg; S M Drance
Journal:  Ophthalmology       Date:  1997-03       Impact factor: 12.079

4.  Reproducibility of topographic measurements of the optic nerve head with laser tomographic scanning.

Authors:  F E Kruse; R O Burk; H E Völcker; G Zinser; U Harbarth
Journal:  Ophthalmology       Date:  1989-09       Impact factor: 12.079

5.  Racial differences in optic nerve head parameters.

Authors:  T Chi; R Ritch; D Stickler; B Pitman; C Tsai; F Y Hsieh
Journal:  Arch Ophthalmol       Date:  1989-06

6.  The area of the neuroretinal rim of the optic nerve in normal eyes.

Authors:  R J Britton; S M Drance; M Schulzer; G R Douglas; D K Mawson
Journal:  Am J Ophthalmol       Date:  1987-04-15       Impact factor: 5.258

7.  Optic disc measurement: a comparison of indirect ophthalmoscopic methods.

Authors:  A F Spencer; S A Vernon
Journal:  Br J Ophthalmol       Date:  1995-10       Impact factor: 4.638

8.  Detection of optic disc change with the Heidelberg retina tomograph before confirmed visual field change in ocular hypertensives converting to early glaucoma.

Authors:  D S Kamal; A C Viswanathan; D F Garway-Heath; R A Hitchings; D Poinoosawmy; C Bunce
Journal:  Br J Ophthalmol       Date:  1999-03       Impact factor: 4.638

9.  Identification of early glaucoma cases with the scanning laser ophthalmoscope.

Authors:  G Wollstein; D F Garway-Heath; R A Hitchings
Journal:  Ophthalmology       Date:  1998-08       Impact factor: 12.079

10.  Correlation of visual field with scanning confocal laser optic disc measurements in glaucoma.

Authors:  L Brigatti; J Caprioli
Journal:  Arch Ophthalmol       Date:  1995-09
View more
  3 in total

1.  Optic disc topographic parameters measured in the normal cynomolgus monkey by confocal scanning laser tomography.

Authors:  T Taniguchi; M Shimazawa; M Araie; G Tomita; M Sasaoka; Y Kitazawa; H Hara
Journal:  Br J Ophthalmol       Date:  2005-08       Impact factor: 4.638

2.  Optic disc cup slope and visual field indices in normal, ocular hypertensive and early glaucomatous eyes.

Authors:  A B Cullinane; A Waldock; J P Diamond; J M Sparrow
Journal:  Br J Ophthalmol       Date:  2002-05       Impact factor: 4.638

3.  The use of the scanning laser ophthalmoscope in the evaluation of amblyopia (an American Ophthalmological Society thesis).

Authors:  David A Johnson
Journal:  Trans Am Ophthalmol Soc       Date:  2006