The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions.

.   

Abstract

OBJECTIVE: To evaluate the sphygmomanometers calibration accuracy and the physical conditions of the cuff-bladder, bulb, pump, and valve.
METHODS: Sixty hundred and forty five aneroid sphygmomanometers were evaluated, 521 used in private practice and 124 used in hospitals. Aneroid manometers were tested against a properly calibrated mercury manometer and were considered calibrated when the error was </=3mm Hg. The physical conditions of the cuffs-bladder, bulb, pump, and valve were also evaluated.
RESULTS: Of the aneroid sphygmomanometers tested, 51% of those used in private practice and 56% of those used in hospitals were found to be not accurately calibrated. Of these, the magnitude of inaccuracy ranged from 4 to 8mm Hg in 70% and 51% of the devices, respectively. The problems found in the cuffs - bladders, bulbs, pumps, and valves of the private practice and hospital devices were bladder damage (34% vs. 21%, respectively), holes/leaks in the bulbs (22% vs. 4%, respectively), and rubber aging (15% vs. 12%, respectively). Of the devices tested, 72% revealed at least one problem interfering with blood pressure measurement accuracy.
CONCLUSION: Most of the manometers evaluated, whether used in private practice or in hospitals, were found to be inaccurate and unreliable, and their use may jeopardize the diagnosis and treatment of arterial hypertension.

Entities:  

Year:  2000        PMID: 10904278

Source DB:  PubMed          Journal:  Arq Bras Cardiol        ISSN: 0066-782X            Impact factor:   2.000


  1 in total

1.  The extent of inaccurate aneroid sphygmomanometers in a hospital setting.

Authors:  Dimitri A Cozanitis; Christopher J Jones
Journal:  Wien Med Wochenschr       Date:  2010-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.