A new verification film system for routine quality control of radiation fields: Kodak EC-L.

A Hermann1, K Bratengeier, A Priske, M Flentje.   

Abstract

BACKGROUND: The use of modern irradiation techniques requires better verification films for determining set-up deviations and patient movements during the course of radiation treatment. This is an investigation of the image quality and time requirement of a new verification film system compared to a conventional portal film system.
MATERIAL AND METHODS: For conventional verifications we used Agfa Curix HT 1000 films which were compared to the new Kodak EC-L film system. 344 Agfa Curix HT 1000 and 381 Kodak EC-L portal films of different tumor sites (prostate, rectum, head and neck) were visually judged on a light box by 2 experienced physicians. Subjective judgement of image quality, masking of films and time requirement were checked.
RESULTS: In this investigation 68% of 175 Kodak EC-L ap/pa-films were judged "good", only 18% were classified "moderate" or "poor" 14%, but only 22% of 173 conventional ap/pa verification films (Agfa Curix HT 1000) were judged to be "good".
CONCLUSIONS: The image quality, detail perception and time required for film inspection of the new Kodak EC-L film system was significantly improved when compared with standard portal films. They could be read more accurately and the detection of set-up deviation was facilitated.

Entities:  

Mesh:

Year:  2000        PMID: 10897254     DOI: 10.1007/s000660050009

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  1 in total

1.  Abstracts of the 1st International Conference on the Clinical Use of Tomotherapy. October 17-18, 2008. Munich, Germany.

Authors: 
Journal:  Strahlenther Onkol       Date:  2009-01       Impact factor: 3.621

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.