M Mazereeuw1, V Spikmans, U R Tjaden, J van der Greef.
Abstract
The use of isotachophoretic (ITP) sample focusing to improve the detection limits for the analysis of charged compounds in capillary electrochromatography (CEC) is described. A coupled-column set-up was used with a 220-microm inner diameter capillary, in which counterflow ITP focusing was performed, connected via a T-junction to a 75-microm inner diameter CEC capillary. As is illustrated, the use of ITP focusing resulted in a dramatic reduction of the sample concentration detection limits. To demonstrate the performance of the ITP-CEC combination, several cationic low-molecular mass compounds in a plasma and urine matrix are analysed using UV-absorbance and mass spectrometric detection. A linear calibration curve was constructed over three decades and detection limits in the low nmol/l range were found for academic samples, using UV-absorbance detection.Mesh:
Year: 2000 PMID: 10893038 DOI: 10.1016/s0021-9673(00)00259-4
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759