| Literature DB >> 10892594 |
T Gordi1, E Nielsen, Z Yu, D Westerlund, M Ashton.
Abstract
A previously established HPLC system with post-column derivatization for the analysis of artemisinin was coupled to an ADS (alkyl-diol silica) pre-column, allowing direct and repetitive injection of protein-rich fluids such as plasma. The limit of quantitation for 100 microl of plasma was 10 ng/ml (CV=10.5%) while concentrations down to 2 ng/ml could be quantified for 1.00 ml saliva samples (CV=11.1%). The system was linear in the tested range of 10-2000 ng/ml for plasma and 2-240 ng/ml for saliva samples, respectively. This paper introduces coupled column HPLC as a simplified method for the routine analysis of artemisinin in biological fluids.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10892594 DOI: 10.1016/s0378-4347(00)00156-0
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273