Abstract
A mixture of saponin compounds (saikosaponin c, a, and d) in the 70% ethanol extract of a powdered sample of Bupleuri radix are analyzed by an Inertsil ODS-3 C(18) column at a flow rate of 1.0 mL/min and detection wavelength of 203 nm. Well resolved chromatograms of saikosaponin c, a, and d are obtained with a gradient elution of acetonitrile-water from 40:60 (v/v) to 50:50 (v/v). The total time required for a single analysis is approximately 20 min. Calibration curves for saikosaponin c, a, and d are linear up to 2.5 mg/mL. The coefficient of variability values for saikosaponins in the extract are below 4%, and the recoveries for saikosaponin c, a, and d are 95.2 +/- 1.1, 96.5 +/- 0.9, and 96.2 +/- 1.0%, respectively. The changes in saikosaponin contents for a two-year growth of Bupleurum falcatum are measured by the established high-performance liquid chromatography method.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10890745 DOI: 10.1093/chromsci/38.6.229
Source DB: PubMed Journal: J Chromatogr Sci ISSN: 0021-9665 Impact factor: 1.618