Literature DB >> 10890597

A new alignment jig for quantification and prescription of three-dimensional alignment for the patellar-tendon-bearing trans-tibial prosthesis.

S W Sin1, D H Chow, J C Cheng.   

Abstract

Clinically, it is hard to achieve and reproduce prosthesis alignment at will during daily prosthesis fitting. A new alignment jig was designed and developed to facilitate quantification and prescription of prosthesis alignment for patellar-tendon-bearing (PTB) trans-tibial prostheses. The alignment jig provided instantaneous readings of the three-dimensional orientation and position of the socket relative to the prosthetic foot in standardised units. The inter- and intra-tester errors of the alignment jig in measuring prosthesis alignment were evaluated and demonstrated to have good reliability. The alignment jig was recommended to be used clinically after the conventional dynamic alignment procedure to document the prosthesis alignment. Further application of the alignment jig for systematic evaluation of the effects of prosthesis alignment on gait for trans-tibial amputees is suggested.

Mesh:

Year:  1999        PMID: 10890597     DOI: 10.3109/03093649909071638

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895


  1 in total

Review 1.  Systematic Review of Studies Examining Transtibial Prosthetic Socket Pressures with Changes in Device Alignment.

Authors:  Philip Davenport; Siamak Noroozi; Philip Sewell; Saeed Zahedi
Journal:  J Med Biol Eng       Date:  2017-01-21       Impact factor: 1.553