| Literature DB >> 10886072 |
J Anderson1, H G Wood, P E Allaire, D B Olsen.
Abstract
The CFVAD3 is the third prototype of a continuous flow ventricular assist device being developed for implantation in humans. The pump consists of a fully shrouded 4-blade impeller supported by magnetic bearings. On either side of this suspended rotating impeller is a small clearance region through which the blood flows. The spacing and geometry of these clearance regions are very important to the successful operation of this blood pump. Computational fluid dynamics (CFD) solutions for this flow were obtained using TascFlow, a software package available from AEA Technology, U.K. Flow in these clearance regions was studied parametrically by varying the size of the clearance, the blood flow rate into the pump, and the rotational speed of the pump. The numerical solutions yield the direction and magnitude of the flow and the dynamic pressure. Experimentally measured pump flow rates are compared to the numerical study. The results of the study provide guidance for improving pump efficiency. It is determined that current clearances can be significantly reduced to improve pump efficiency without negative impacts.Entities:
Mesh:
Year: 2000 PMID: 10886072 DOI: 10.1046/j.1525-1594.2000.06580.x
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094