[Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia].

M Okuyama1, I Nakamura, O Kemmotsu.   

Abstract

We describe our experience of an accidental decrease in the air flow during air/oxygen sevoflurane anesthesia. According to the malfunction of an air compressor system, dehydration of the compressed air was inadequate, and the wet air was delivered to the anesthesia machine. The moisture was formed at a flow control valve to cause a gradual decrease in the air flow. The low-pressure alarm did not work at that time, because the pressure of the air supply was normal. There are two types of medical air, one is produced from the ambient air by air compressor system and another is the synthetic air which is a mixture of oxygen and nitrogen. Although the synthetic air is always dry and clean, the former contains dust, bacteria, and moisture. There is a possibility that all of these particles cannot be excluded before use. Accordingly, we must check the air compressor system routinely when a source of the compressed air is used.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10885247

Source DB:  PubMed          Journal:  Masui        ISSN: 0021-4892


  3 in total

1.  Sudden Appearance of Water in Flowmeter During Air/Oxygen and Sevoflurane Anaesthesia.

Authors:  Tünay Kandemir; Selda Muslu; Erbin Kandemir
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-09-09

2.  Critical incident is a possibility with water in flowmeter.

Authors:  M Manjuladevi; Ks Vasudeva Upadhyaya; Ps Sathyanarayana; Rashmi Rani; Viraj Shah
Journal:  Indian J Anaesth       Date:  2014 Nov-Dec

3.  Water contamination of the central air supply: Out of sight, out of mind!

Authors:  Geeta Kamal; Anju Gupta; Sapna Bathla; Aikta Gupta
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2018 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.