Abstract
A new germanium spectrometer especially suited for large sample measurements is described in detail. It is operated in the Gran Sasso underground laboratory under shielding rock of 3300 m water equivalent, which reduces the muon flux by six orders of magnitude. The integral background counting rate in the energy range from 50 to 2750 keV is about 0.15 min(-1). The low peak count rates of mostly less than 1 count per day together with a relative efficiency of 102% and the high sample capacity makes this spectrometer one of the most sensitive worldwide. Some sample measurements for the solar neutrino experiment BOREXINO and the detector efficiency calibration by the Monte Carlo method are discussed as well.Entities:
Year: 2000 PMID: 10879860 DOI: 10.1016/s0969-8043(00)00132-9
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513