Optimising the performance of intermittent pneumatic compression devices.

N Labropoulos1, J Cunningham, S S Kang, M A Mansour, W H Baker.   

Abstract

OBJECTIVES: this study was designed to determine whether an intermittent pneumatic compression device (IPC) with an increased maximal inflation pressure, a decreased time to maximal pressure and a longer duration of compression would improve venous return compared to a standard IPC device.
METHODS: thirty limbs in 15 volunteers without evidence of venous disease were studied using duplex scanning at rest and during the application of two different IPC devices with different compression parameters. The first device IPC-1 (SCD 5325, Kendall) has a six-chambered cuff applying 45 mmHg after 12 s, sequentially from ankle to thigh followed by 60 s of non-compression. The second device IPC-2 (Vena-Assist(R), ACI Medical) has a foot, ankle and calf cuff, applies a pressure of 80 mmHg, has a pressure rise time of 0.3 s, maintains inflation for 5.5 s, and has a cycling time of 1 min. Peak venous velocity and acceleration time were measured at rest and during the IPC application. Measurements were obtained in supine position from the common femoral vein 1 cm above the saphenofemoral junction to include the entire venous outflow from the limb.
RESULTS: peak venous velocity at rest was significantly higher in the right limb than in the left limb (26+/-7.2 vs. 22+/-5.7 cm/s, p<0.01). Peak venous velocity was significantly increased by both IPC devices (p <0.0001). IPC-2 achieved significantly higher peak venous velocity than IPC-1 (55.1+/-17.8 vs. 37.4+/-6.9 cm/s, p<0.0001). Acceleration time was also found to be significantly shorter (370+/-93.4 vs. 560+/-83.5 ms, p<0.0001) in IPC-2 than in IPC-1, respectively.
CONCLUSIONS: we have demonstrated that progressive inflation at the foot, ankle and calf, increasing maximal inflation pressure and decreasing time to maximal pressure result in increased venous return. These changes may improve the efficacy of IPC devices in the prevention of deep-venous thrombosis (DVT) formation. Copyright 2000 Harcourt Publishers Ltd.

Entities:  

Mesh:

Year:  2000        PMID: 10873726     DOI: 10.1053/ejvs.2000.1067

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  2 in total

1.  Enhanced muscle blood flow with intermittent pneumatic compression of the lower leg during plantar flexion exercise and recovery.

Authors:  K A Zuj; C N Prince; R L Hughson; S D Peterson
Journal:  J Appl Physiol (1985)       Date:  2017-11-09

2.  Thromboelastographic changes during laparoscopic fundoplication.

Authors:  Indre Zostautiene; Kristina Zvinienė; Darius Trepenaitis; Rolandas Gerbutavičius; Antanas Mickevičius; Rima Gerbutavičienė; Mindaugas Kiudelis
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2017-03-13       Impact factor: 1.195

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.