Fatigue failure of an AO spiral blade.

A A Syed1, J G Kennedy, H Mullet, J O'Flanagan, D Taylor.   

Abstract

We report an unusual case of a femoral neck stress fracture leading to the fatigue failure of an AO spiral blade. An unreamed femoral nail with a spiral blade was inserted to treat an unstable subtrochanteric femoral fracture. which lead to fracture union at 5 months. Eight months post-operatively the patient started to complain of left hip pain. Serial radiographs revealed progressive osteoporosis of the proximal femur possibly due to the stress sharing effect of a stiff intramedullary device, which continued to bear a significant amount of the transmitted load. The cause of pain was a stress fracture of the femoral neck and the AO spiral blade, which only became radiologically visible 4 months after the start of the symptoms (1 year after the initial operation). The implant was removed and replaced by a cemented hemiarthroplasty. This case reaffirms the difficulty in diagnosing a stress fracture through a metallic implant. The delay in diagnosis may be shortened if stress fracture were included as an expected complication following an intramedullary nailing.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10853918     DOI: 10.1007/s004020050487

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  2 in total

Review 1.  Fixation principles in metaphyseal bone--a patent based review.

Authors:  R Curtis; J Goldhahn; R Schwyn; P Regazzoni; N Suhm
Journal:  Osteoporos Int       Date:  2004-11-03       Impact factor: 4.507

2.  Symptoms indicating imminent breakage of a femoral interlocking nail: a case report.

Authors:  Kl Pan; Wh Chan
Journal:  Malays Orthop J       Date:  2013-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.