Efficacy of transverse tripolar stimulation for relief of chronic low back pain: results of a single center.

K V Slavin1, K J Burchiel, V C Anderson, B Cooke.   

Abstract

The goal of this study was to evaluate the efficacy of the transverse tripolar spinal cord stimulation system (TTS) in providing relief of low back pain in patients with chronic non-malignant pain. Transverse tripolar electrodes were implanted in the lower thoracic region (T(8-9) to T(12)-L(1)) in 10 patients with chronic neuropathic pain, all of whom reported a significant component of low back pain in combination with unilateral or bilateral leg pain. One patient reported inadequate pain relief during the trial and was not implanted with a permanent generator. A visual analogue scale of low back pain showed a nonsignificant decrease from 64 +/- 19 to 47 +/- 30 (p = 0.25; paired t test) after 1 month of stimulation. Similarly, functional disability evaluated using Oswestry Low Back Pain Questionnaire was not improved (p = 0. 46; paired t test). We conclude that chronic low back pain is not particularly responsive to the transverse stimulation provided by the TTS system. Copyright 2000 S. Karger AG, Basel

Entities:  

Mesh:

Year:  1999        PMID: 10853117     DOI: 10.1159/000029770

Source DB:  PubMed          Journal:  Stereotact Funct Neurosurg        ISSN: 1011-6125            Impact factor:   1.875