Peak expiratory flow profiles delivered by pump systems. Limitations due to wave action.

M R Miller1, B Jones, Y Xu, O F Pedersen, P H Quanjer.   

Abstract

Pump systems are currently used to test the performance of both spirometers and peak expiratory flow (PEF) meters, but for certain flow profiles the input signal (i.e., requested profile) and the output profile can differ. We developed a mathematical model of wave action within a pump and compared the recorded flow profiles with both the input profiles and the output predicted by the model. Three American Thoracic Society (ATS) flow profiles and four artificial flow-versus-time profiles were delivered by a pump, first to a pneumotachograph (PT) on its own, then to the PT with a 32-cm upstream extension tube (which would favor wave action), and lastly with the PT in series with and immediately downstream to a mini-Wright peak flow meter. With the PT on its own, recorded flow for the seven profiles was 2.4 +/- 1.9% (mean +/- SD) higher than the pump's input flow, and similarly was 2.3 +/- 2.3% higher than the pump's output flow as predicted by the model. With the extension tube in place, the recorded flow was 6.6 +/- 6.4% higher than the input flow (range: 0.1 to 18.4%), but was only 1.2 +/- 2.5% higher than the output flow predicted by the model (range: -0.8 to 5.2%). With the mini-Wright meter in series, the flow recorded by the PT was on average 6.1 +/- 9.1% below the input flow (range: -23.8 to 2. 5%), but was only 0.6 +/- 3.3% above the pump's output flow predicted by the model (range: -5.5 to 3.9%). The mini-Wright meter's reading (corrected for its nonlinearity) was on average 1.3 +/- 3.6% below the model's predicted output flow (range: -9.0 to 1. 5%). The mini-Wright meter would be deemed outside ATS limits for accuracy for three of the seven profiles when compared with the pump's input PEF, but this would be true for only one profile when compared with the pump's output PEF as predicted by the model. Our study shows that the output flow from pump systems can differ from the input waveform depending on the operating configuration. This effect can be predicted with reasonable accuracy using a model based on nonsteady flow analysis that takes account of pressure wave reflections within pump systems.

Mesh:

Year:  2000        PMID: 10852762     DOI: 10.1164/ajrccm.161.6.9805085

Source DB:  PubMed          Journal:  Am J Respir Crit Care Med        ISSN: 1073-449X            Impact factor:   21.405


  2 in total

1.  Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device.

Authors:  M R Miller; P R Atkins; O F Pedersen
Journal:  Thorax       Date:  2003-05       Impact factor: 9.139

2.  Can a normal peak expiratory flow exclude severe chronic obstructive pulmonary disease?

Authors:  R Perez-Padilla; W M Vollmer; J C Vázquez-García; P L Enright; A M B Menezes; A S Buist
Journal:  Int J Tuberc Lung Dis       Date:  2009-03       Impact factor: 2.373

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.