Literature DB >> 10851106

Biomechanical assessment of transoral plate fixation for atlantoaxial instability.

F Kandziora1, F Kerschbaumer, M Starker, T Mittlmeier.   

Abstract

STUDY
DESIGN: In an experimental study using human cadaver specimens the biomechanical data of anterior atlantoaxial plating according to Harms were evaluated.
OBJECTIVES: The purpose of this study was to evaluate this method biomechanically. SUMMARY AND BACKGROUND DATA: The optimum fixation method to achieve fusion at the atlantoaxial joint after odontoid resection is still a matter of discussion. Isolated posterior surgical procedures for treatment of irreducible atlantoaxial kyphosis with spinal cord compression are associated with high rates of morbidity and mortality. Transoral atlantoaxial plate fixation has been designed by Harms as a fixation technique after odontoid resection. In a modification, this procedure has been combined with the posterior wire fusion according to Brooks.
METHOD: Eight human cadaver craniocervical specimens were tested in flexion, extension, rotation, and bending with a nondestructive flexibility method using a nonconstrained testing apparatus. Five different groups were examined: 1) control group (intact); 2) unstable group (after dissection of the atlantoaxial ligaments and odontoidectomy), 3) Harms group (transoral atlantoaxial plate fixation) 4) Harms-Brooks group (transoral atlantoaxial plate fixation and dorsal atlantoaxial wire fixation); and 5) Magerl group (transarticular atlantoaxial screw fixation). In a second experimental series, failure loads of the Harms-Brooks and the Magerl fixation methods were determined.
RESULTS: The angular displacement of the Harms-Brooks group and the Magerl group was less than in any other group. Stiffness values at 0-3.0 Nm loads in any direction were larger for the Harms-Brooks-and Magerl-fixated specimens than for the Harms, control, or unstable specimens. No statistically significant difference was observed between Harms-Brooks and Magerl reconstruction stiffness. Ultimate failure load in the Harms-Brooks group was higher than in the Magerl group.
CONCLUSIONS: Experimentally, isolated anterior atlantoaxial plating was less stable than the combined reconstruction procedures. Transoral plate fixation according to Harms in combination with posterior wire fixation according to Brooks provided a failure load and stiffness equal to transarticular screw fixation according to Magerl.

Entities:  

Mesh:

Year:  2000        PMID: 10851106     DOI: 10.1097/00007632-200006150-00016

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  14 in total

Review 1.  [Vertebral body replacement in spine surgery].

Authors:  F Kandziora; K J Schnake; C K Klostermann; N P Haas
Journal:  Unfallchirurg       Date:  2004-05       Impact factor: 1.000

2.  [Expandable cages: biomechanical comparison of different cages for ventral spondylodesis in the thoracolumbar spine].

Authors:  C Khodadadyan-Klostermann; J Schaefer; Ph Schleicher; R Pflugmacher; T Eindorf; N P Haas; F Kandziora
Journal:  Chirurg       Date:  2004-07       Impact factor: 0.955

3.  Dose-dependent effects of combined IGF-I and TGF-beta1 application in a sheep cervical spine fusion model.

Authors:  F Kandziora; R Pflugmacher; M Scholz; J Schäfer; G Schollmeier; G Schmidmaier; G Duda; M Raschke; N P Haas
Journal:  Eur Spine J       Date:  2002-11-08       Impact factor: 3.134

4.  Atlantoaxial fusion using anterior transarticular screw fixation of C1-C2: technical innovation and biomechanical study.

Authors:  Milan K Sen; Thomas Steffen; Lorne Beckman; Anthony Tsantrizos; Rudolph Reindl; Max Aebi
Journal:  Eur Spine J       Date:  2005-01-25       Impact factor: 3.134

5.  Anterior retropharyngeal fixation C1-2 for stabilization of atlantoaxial instabilities: study of feasibility, technical description and preliminary results.

Authors:  Heiko Koller; Volker Kammermeier; Dietmar Ulbricht; Allan Assuncao; Stefan Karolus; Boris van den Berg; Ulrich Holz
Journal:  Eur Spine J       Date:  2006-04-08       Impact factor: 3.134

Review 6.  [Instability of the upper cervical spine due to rheumatism].

Authors:  C E Heyde; U Weber; R Kayser
Journal:  Orthopade       Date:  2006-03       Impact factor: 1.087

7.  C1-C2 arthrodesis after transoral odontoidectomy and suboccipital craniectomy for ventral brain stem compression in Chiari I patients.

Authors:  Steven W Hwang; Carl B Heilman; Ron I Riesenburger; James Kryzanski
Journal:  Eur Spine J       Date:  2008-07-16       Impact factor: 3.134

8.  C1-2 transarticular screws combined with C1 laminar hooks fixation: a modified posterior atlantoaxial fixation technique and outcome in 72 patients.

Authors:  Bin Ni; Xiang Guo; Ning Xie; Songkai Li; Fengjing Zhou; Feng Zhang; Qi Liu
Journal:  Eur Spine J       Date:  2012-08-21       Impact factor: 3.134

Review 9.  [Therapy of cervical rheumatoid arthritis].

Authors:  R Kothe; L Wiesner; W Rüther
Journal:  Z Rheumatol       Date:  2004-08       Impact factor: 1.372

10.  Biomechanical evaluation of four different posterior screw and rod fixation techniques for the treatment of the odontoid fractures.

Authors:  Lei Li; Wen-Fei Liu; Hong-Kun Jiang; Yun-Peng Li
Journal:  Int J Clin Exp Med       Date:  2015-06-15
View more