Literature DB >> 10848679

Computational flow study of the continuous flow ventricular assist device, prototype number 3 blood pump.

J B Anderson1, H G Wood, P E Allaire, G Bearnson, P Khanwilkar.   

Abstract

A computational fluid dynamics study of blood flow in the continuous flow ventricular assist device, Prototype No. 3 (CFVAD3), which consists of a 4 blade shrouded impeller fully supported in magnetic bearings, was performed. This study focused on the regions within the pump where return flow occurs to the pump inlet, and where potentially damaging shear stresses and flow stagnation might occur: the impeller blade passages and the narrow gap clearance regions between the impeller-rotor and pump housing. Two separate geometry models define the spacing between the pump housing and the impeller's hub and shroud, and a third geometry model defines the pump's impeller and curved blades. The flow fields in these regions were calculated for various operating conditions of the pump. Pump performance curves were calculated, which compare well with experimentally obtained data. For all pump operating conditions, the flow rates within the gap regions were predicted to be toward the inlet of the pump, thus recirculating a portion of the impeller flow. Two smaller gap clearance regions were numerically examined to reduce the recirculation and to improve pump efficiency. The computational and geometry models will be used in future studies of a smaller pump to determine increased pump efficiency and the risk of hemolysis due to shear stress, and to insure the washing of blood through the clearance regions to prevent thrombosis.

Entities:  

Mesh:

Year:  2000        PMID: 10848679     DOI: 10.1046/j.1525-1594.2000.06442.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  4 in total

Review 1.  The use of computational fluid dynamics in the development of ventricular assist devices.

Authors:  Katharine H Fraser; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Eng Phys       Date:  2010-11-13       Impact factor: 2.242

2.  Evaluation of a pumping assist lung that uses a rotating fiber bundle.

Authors:  Robert G Svitek; Brian J Frankowski; William J Federspiel
Journal:  ASAIO J       Date:  2005 Nov-Dec       Impact factor: 2.872

3.  Computational fluid dynamics analysis of blade tip clearances on hemodynamic performance and blood damage in a centrifugal ventricular assist device.

Authors:  Jingchun Wu; Bradley E Paden; Harvey S Borovetz; James F Antaki
Journal:  Artif Organs       Date:  2009-10-12       Impact factor: 3.094

4.  Comparison and experimental validation of fluid dynamic numerical models for a clinical ventricular assist device.

Authors:  Jiafeng Zhang; Pei Zhang; Katharine H Fraser; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2013-02-27       Impact factor: 3.094