Literature DB >> 10837962

Open magnetic resonance imaging using titanium-zirconium needles: improved accuracy for interstitial brachytherapy implants?

Y Popowski1, E Hiltbrand, D Joliat, M Rouzaud.   

Abstract

PURPOSE: To evaluate the benefit of using an open magnetic resonance (MR) machine and new MR-compatible needles to improve the accuracy of brachytherapy implants in pelvic tumors. METHODS AND MATERIALS: The open MR machine, foreseen for interventional procedures, allows direct visualization of the pelvic structures that are to be implanted. For that purpose, we have developed MR- and CT-compatible titanium-zirconium (Ti-Zr) brachytherapy needles that allow implantations to be carried out under the magnetic field. In order to test the technical feasibility of this new approach, stainless steel (SS) and Ti-Zr needles were first compared in a tissue-equivalent phantom. In a second step, two patients implanted with Ti-Zr needles in the brachytherapy operating room were scanned in the open MR machine. In a third phase, four patients were implanted directly under open MR control.
RESULTS: The artifacts induced by both materials were significantly different, strongly favoring the Ti-Zr needles. The implantation in both first patients confirmed the excellent quality of the pictures obtained with the needles in vivo and showed suboptimal implant geometry in both patients. In the next 4 patients, the tumor could be punctured with excellent accuracy, and the adjacent structures could be easily avoided.
CONCLUSION: We conclude that open MR using MR-compatible needles is a very promising tool in brachytherapy, especially for pelvic tumors.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10837962     DOI: 10.1016/s0360-3016(00)00488-0

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

Review 1.  Research and development of metals for medical devices based on clinical needs.

Authors:  Takao Hanawa
Journal:  Sci Technol Adv Mater       Date:  2012-12-13       Impact factor: 8.090

2.  Venezia applicator with oblique needles improves clinical target volume coverage in distal parametrial tumor residue compared to parallel needles only.

Authors:  Manon Kissel; Nathalie Fournier-Bidoz; Olivier Henry; Sophie Bockel; Tamizhanban Kumar; Sophie Espenel; Cyrus Chargari
Journal:  J Contemp Brachytherapy       Date:  2021-02-18

Review 3.  MR fluoroscopy in vascular and cardiac interventions (review).

Authors:  Maythem Saeed; Steve W Hetts; Joey English; Mark Wilson
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-26       Impact factor: 2.357

Review 4.  Feasibility of functional imaging for brachytherapy.

Authors:  Alfredo Polo
Journal:  J Contemp Brachytherapy       Date:  2009-03-23

5.  Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (IV): Basic principles and parameters for MR imaging within the frame of image based adaptive cervix cancer brachytherapy.

Authors:  Johannes C A Dimopoulos; Peter Petrow; Kari Tanderup; Primoz Petric; Daniel Berger; Christian Kirisits; Erik M Pedersen; Erik van Limbergen; Christine Haie-Meder; Richard Pötter
Journal:  Radiother Oncol       Date:  2012-01-30       Impact factor: 6.280

  5 in total