| Literature DB >> 10836429 |
C Fainstein1, E Winkler, M Saravi.
Abstract
We report Alanine Dosimeter preparation, procedures for using the ESR/Dosimetry method, and the resulting calibration curve for gamma-irradiation in the range from 10-30 Gy. We use calibration curve to measure the irradiation dose in gamma-irradiation of human blood, as required in Blood Transfusion Therapy. The ESR/Alanine results are compared against those obtained using the thermoluminescent dosimetry (TLD) method.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10836429 DOI: 10.1016/s0969-8043(00)00070-1