Detection of subdural placement of epidural catheter using nerve stimulation.

B C Tsui1, S Gupta, D Emery, B Finucane.   

Abstract

PURPOSE: To report the detection of a subdural catheter placement using nerve stimulation through an epidural catheter. CLINICAL FEATURES: An 85-yr-old gentleman was scheduled for radical cystectomy and creation of an ileal conduit. Combined general anesthesia and regional technqiue was selected. An epidural catheter (19 G Arrow Flextip Plus) was inserted prior to induction of general anesthesia. Intra-operatively, the patient received 5 mg morphine and 10 ml bupivacane 0.5% via the epidural catheter. The patient remained hemodynamically stable throughout the operation and did not require intravenous opioids. The patient was discharged to the ward with an order for epidural morphine for pain control. The next day, the patient remained comfortable. As an ongoing quality assessment to survey the success rate of epidural catheters at our institution, all patients are invited to have their catheter assessed using an electrical epidural stimulation test. Electrical stimulation (1-10 mA) with a segmental motor response (truncal or extremities movement) indicates that the catheter is in the epidural space. No motor response indicates that it is not. In this case, subdural catheter placement was suspected because a diffuse motor response including right anterior chest wall, back muscle, and bilateral lower extremities was observed using only 0.3 mA. Subdural catheter placement was subsequently confirmed by a radiograph showing a very thin film of dye spreading cephalad and caudad over many segments.
CONCLUSION: This new electrical test helps to detect subdural placement objectively.

Entities:  

Mesh:

Year:  2000        PMID: 10831207     DOI: 10.1007/BF03018980

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  3 in total

Review 1.  Innovative approaches to neuraxial blockade in children: the introduction of epidural nerve root stimulation and ultrasound guidance for epidural catheter placement.

Authors:  Ban C H Tsui
Journal:  Pain Res Manag       Date:  2006       Impact factor: 3.037

Review 2.  Localization of epidural space: A review of available technologies.

Authors:  Hesham Elsharkawy; Abraham Sonny; Ki Jinn Chin
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Jan-Mar

3.  Minimum current requirement for confirming the localization of an epiradicular catheter placement.

Authors:  Ji Seon Jeong; Jae Chol Shim; Jae Hang Shim; Dong Won Kim; Min Serk Kang
Journal:  Korean J Anesthesiol       Date:  2012-09-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.