Literature DB >> 10823106

Monitoring equipment induced tachycardia in patients with minute ventilation rate-responsive pacemakers.

P A Southorn1, G S Kamath, G M Vasdev, D L Hayes.   

Abstract

Minute ventilation-sensing pacemakers enable the paced heart to respond to an increased workload. Two patients with such a pacemaker developed pacemaker-driven tachycardia when connected to an electrocardiogram (ECG) monitor also capable of documenting ventilatory frequency and ECG lead disconnection. This tachycardia stopped when the ECG leads were removed. These pacemakers and monitors emit a low-amplitude electrical current and measure the resultant impedence signal across the chest. When patients are connected to the monitor the pacemaker sensor summates both impedence signals and the paced heart rate is increased as a result.

Entities:  

Mesh:

Year:  2000        PMID: 10823106     DOI: 10.1093/oxfordjournals.bja.a013480

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

2.  A patient with dizziness, tachycardia and a DDDR pacemaker.

Authors:  J C Balt; P Dekker; W G de Voogt
Journal:  Neth Heart J       Date:  2006-01       Impact factor: 2.380

3.  Echocardiography induced tachycardia in a patient with a minute ventilation rate responsive pacemaker.

Authors:  Chotikorn Khunnawat; Siddharth Mukerji; Sundar Sankaran; Ranjan K Thakur
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

4.  Sensors for rate responsive pacing.

Authors:  Simonetta Dell'Orto; Paolo Valli; Enrico Maria Greco
Journal:  Indian Pacing Electrophysiol J       Date:  2004-07-01

5.  Pacemaker-Mediated Tachycardia: A Case Report.

Authors:  Rajwinder Gill; Vineet Meghrajani
Journal:  Cureus       Date:  2022-07-05
  5 in total