Skin damage probabilities using fixation materials in high-energy photon beams.

J Carl1, A Vestergaard.   

Abstract

INTRODUCTION: Patient fixation, such as thermoplastic masks, carbon-fibre support plates and polystyrene bead vacuum cradles, is used to reproduce patient positioning in radiotherapy. Consequently low-density materials may be introduced in high-energy photon beams. The aim of the this study was to measure the increase in skin dose when low-density materials are present and calculate the radiobiological consequences in terms of probabilities of early and late skin damage.
METHOD: An experimental thin-windowed plane-parallel ion chamber was used. Skin doses were measured using various overlaying low-density fixation materials. A fixed geometry of a 10x10 cm field, a SSD=100 cm and photon energies of 4, 6 and 10 MV on Varian Clinac 2100C accelerators were used for all measurements. Radiobiological consequences of introducing these materials into the high-energy photon beams were evaluated in terms of early and late damage of the skin based on the measured surface doses and the LQ-model.
RESULTS: The experimental ion chamber gave results consistent with other studies. A relationship between skin dose and material thickness in mg/cm(2) was established and used to calculate skin doses in scenarios assuming radiotherapy treatment with opposed fields.
CONCLUSION: Conventional radiotherapy may apply mid-point doses up to 60-66 Gy in daily 2-Gy fractions opposed fields. Using thermoplastic fixation and high-energy photons as low as 4 MV do increase the dose to the skin considerably. However, using thermoplastic materials with thickness less than 100 mg/cm(2) skin doses are comparable with those produced by variation in source to skin distance, field size or blocking trays within clinical treatment set-ups. The use of polystyrene cradles and carbon-fibre materials with thickness less than 100 mg/cm(2) should be avoided at 4 MV at doses above 54-60 Gy.

Entities:  

Mesh:

Year:  2000        PMID: 10799732     DOI: 10.1016/s0167-8140(00)00177-8

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  Comparison of surface dose delivered by 7 MV-unflattened and 6 MV-flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

2.  Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.

Authors:  Jürgen Wilbert; Matthias Guckenberger; Bülent Polat; Otto Sauer; Michael Vogele; Michael Flentje; Reinhart A Sweeney
Journal:  Radiat Oncol       Date:  2010-05-26       Impact factor: 3.481

3.  Assessment and Treatment Outcomes of Persistent Radiation-Induced Alopecia in Patients With Cancer.

Authors:  Gregory S Phillips; Morgan E Freret; Danielle Novetsky Friedman; Sabrina Trelles; Oluwaseun Kukoyi; Azael Freites-Martinez; Robin H Unger; Joseph J Disa; Leonard H Wexler; Christopher L Tinkle; James G Mechalakos; Stephen W Dusza; Kathryn Beal; Suzanne L Wolden; Mario E Lacouture
Journal:  JAMA Dermatol       Date:  2020-09-01       Impact factor: 10.282

4.  Reduction of Potential Risk for Skin Toxicity in Megavoltage Radiotherapy Using a Novel Rigid Couch.

Authors:  Mikoto Tamura; Hajime Monzen; Kenji Matsumoto; Masahiko Okumura; Hiroshi Doi; Yasumasa Nishimura
Journal:  In Vivo       Date:  2018 May-Jun       Impact factor: 2.155

5.  An investigation of the depth dose in the build-up region, and surface dose for a 6-MV therapeutic photon beam: Monte Carlo simulation and measurements.

Authors:  Lukkana Apipunyasopon; Somyot Srisatit; Nakorn Phaisangittisakul
Journal:  J Radiat Res       Date:  2012-10-26       Impact factor: 2.724

6.  Surface dose measurements and comparison of unflattened and flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj; Girigesh Yadav; Ananda Giribabu; Karthikeyan Srinivasan; Venkadamanickam Gurusamy; Kothanda Raman; Kaviarasu Karunakaran; Rajesh Thiyagarajan
Journal:  J Med Phys       Date:  2016 Apr-Jun

7.  Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.

Authors:  Eric E Klein; Jacqueline Esthappan; Zuofeng Li
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

8.  Skin dose during radiotherapy: a summary and general estimation technique.

Authors:  Stephen F Kry; Susan A Smith; Rita Weathers; Marilyn Stovall
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

9.  Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop.

Authors:  W Kenji Myint; Malgorzata Niedbala; David Wilkins; Lee H Gerig
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

10.  Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams.

Authors:  Jason Cashmore
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.