A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy.

M Bambynek1, D Flühs, U Quast, D Wegener, C G Soares.   

Abstract

A fast dosimetry system based on plastic scintillator detectors has been developed which allows three-dimensional measurement of the radiation field in water of beta-sources appropriate for application in cardiovascular brachytherapy. This system fulfills the AAPM Task Group 60 recommendations for dosimetry of cardiovascular brachytherapy sources. To demonstrate the use of the system, measurements have been performed with an 90Y-wire source. The dose distribution was determined with a spatial resolution of better than 0.2 mm, with only a few minutes needed per scan. The scintillator dosemeter was absolutely calibrated in terms of absorbed dose to water with a precision of +/-7.5%. The relative precision achievable is +/-2.5%. The response of the system is linear within +/-2% for dose rates from 0.5 mGy s(-1) to 500 mGy s(-1).

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10798687     DOI: 10.1118/1.598927

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

Review 1.  Endovascular brachytherapy--treatment planning and radiation protection.

Authors:  U Quast; D Flühs; M Bambynek
Journal:  Herz       Date:  1998-09       Impact factor: 1.443

2.  Simulation of the precision limits of plastic scintillation detectors using optimal component selection.

Authors:  Frédéric Lacroix; Luc Beaulieu; Louis Archambault; A Sam Beddar
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

Review 3.  Air kerma and absorbed dose standards for reference dosimetry in brachytherapy.

Authors:  T Sander
Journal:  Br J Radiol       Date:  2014-07-07       Impact factor: 3.039

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.