Literature DB >> 10795996

Calibration of a tomotherapeutic MVCT system.

K J Ruchala1, G H Olivera, E A Schloesser, R Hinderer, T R Mackie.   

Abstract

Megavoltage CT provides the ability to image the patient before, during or after a radiotherapy treatment. This allows one to verify not only the placement of a patient's external boundary, but also the locations of internal anatomy. In addition, the reconstructed MVCT values are potentially useful for treatment planning inhomogeneity corrections and dose reconstruction. To this end, dosimetric calibration of the University of Wisconsin Tomotherapy Benchtop MVCT system was investigated. It was found that MVCT values correlate extremely well with electron density and that unlike kilovoltage CT, this correlation is well maintained for higher atomic number materials. Improvements of the order of 1% in the dosimetric calculations of high atomic number materials should be possible by deriving input images from MVCT as opposed to kVCT, and calibrating in terms of electron density, as opposed to physical density.

Entities:  

Mesh:

Year:  2000        PMID: 10795996     DOI: 10.1088/0031-9155/45/4/404

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Improved accuracy of relative electron density and proton stopping power ratio through CycleGAN machine learning.

Authors:  Jessica Scholey; Luciano Vinas; Vasant Kearney; Sue Yom; Peder Eric Zufall Larson; Martina Descovich; Atchar Sudhyadhom
Journal:  Phys Med Biol       Date:  2022-05-02       Impact factor: 4.174

2.  Magnetic resonance imaging for adaptive cobalt tomotherapy: A proposal.

Authors:  Tomas Kron; David Eyles; L John Schreiner; Jerry Battista
Journal:  J Med Phys       Date:  2006-10

3.  The effect and stability of MVCT images on adaptive TomoTherapy.

Authors:  Poonam Yadav; Ranjini Tolakanahalli; Yi Rong; Bhudatt R Paliwal
Journal:  J Appl Clin Med Phys       Date:  2010-07-02       Impact factor: 2.102

4.  Improving dose calculations on tomotherapy MVCT images.

Authors:  Frederik Crop; Antoine Bernard; Nick Reynaert
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

5.  On the molecular relationship between Hounsfield Unit (HU), mass density, and electron density in computed tomography (CT).

Authors:  Atchar Sudhyadhom
Journal:  PLoS One       Date:  2020-12-31       Impact factor: 3.240

6.  An investigation into factors affecting electron density calibration for a megavoltage cone-beam CT system.

Authors:  Jessica Hughes; Lois C Holloway; Alexandra Quinn; Andrew Fielding
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.