Optimized MR imaging for polyacrylamide gel dosimetry.

I C Baustert1, M Oldham, T A Smith, C Hayes, S Webb, M O Leach.   

Abstract

Polyacrylamide gels are a powerful tool to measure radiation dose by quantifying the NMR T2 relaxation times of the irradiated gel. The exploitation of these radiation sensitive gels in clinical radiotherapy requires accurate mapping of T2 values. This paper describes the optimization strategy used to identify accurate and practical methods of measuring the range of T2 values typical of gel dosimeters (140-700 ms). The MR imaging techniques used to measure T2 values and the choice of image acquisition parameters are described. Four sequences are compared and the results are analysed in terms of accuracy, signal-to-noise ratio and acquisition time. A multiple spin echo sequence was found to yield the most accurate results (98.9%). Single spin echo sequences, such as Hahn spin echo and EPI spin echo, were found to measure gel T2 values with an accuracy of 90.1%. This paper reports the importance of careful selection and optimization of the MR imaging sequences for accurate and reliable polyacrylamide gel dosimetry.

Entities:  

Mesh:

Year:  2000        PMID: 10795976     DOI: 10.1088/0031-9155/45/4/303

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Optical-CT gel-dosimetry I: basic investigations.

Authors:  Mark Oldham; Jeffrey H Siewerdsen; Sai Kumar; John Wong; David A Jaffray
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

2.  A practical three-dimensional dosimetry system for radiation therapy.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

3.  A variable echo-number method for estimating R2 in MRI-based polymer gel dosimetry.

Authors:  Yoichi Watanabe; Hitoshi Kubo
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

Review 4.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609