In vitro evaluation of the strength of the conical implant-to-abutment joint in two commercially available implant systems.

M R Norton1.   

Abstract

STATEMENT OF PROBLEM: The cone-screw abutment has been shown to diminish micromovement, reducing the burden of component loosening and fracture. However, it is unclear whether the conical taper and joint design influence strength of the interface, with respect to unfavorable bending moments.
PURPOSE: This comparative study evaluated the resistance to bending for the ITI Straumann and Astra Tech ST implant systems using an 8- and 11-degree internal cone, respectively.
MATERIAL AND METHODS: Assembled units from each system were mounted in a 3-point bending apparatus. High load tests were performed, 4 mm from the joint, and bending moments necessary to induce first point of plastic deformation and ultimate failure were measured. All units were inspected to determine the critical zone of failure.
RESULTS: Bending moments necessary to induce first point of plastic deformation were considered well above that expected in clinical function for both systems. However, the critical zones of failure differed in that the solid Astra abutment deformed before the cone joint with its 11-degree taper and smooth transition into the neck of the screw, preventing screw fracture. By contrast, all ITI screws fractured at the head of the screw where it met the base of the 8-degree cone. It is unclear which aspects of the joint design were responsible for the difference observed in mode of failure or if it was a direct result of the experimental design.
CONCLUSION: For clinically relevant levels of bending moment, no problems were anticipated with respect to component failure for either system.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10793390     DOI: 10.1016/s0022-3913(00)70016-3

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  4 in total

1.  Marginal bone response of implants with platform switching and non-platform switching abutments in posterior healed sites: a 1-year prospective study.

Authors:  Yun-Chi Wang; Joseph Y K Kan; Kitichai Rungcharassaeng; Phillip Roe; Jaime L Lozada
Journal:  Clin Oral Implants Res       Date:  2014-01-03       Impact factor: 5.977

2.  Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading.

Authors:  Mai Ahmed Yousry El-Sheikh; Tamer Mohamed Nasr Mostafa; Mohamed Maamoun El-Sheikh
Journal:  Int J Implant Dent       Date:  2018-12-03

Review 3.  Implant Bio-mechanics for Successful Implant Therapy: A Systematic Review.

Authors:  Khaled Mosfer Alzahrani
Journal:  J Int Soc Prev Community Dent       Date:  2020-11-24

4.  Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.

Authors:  Saied Nokar; Habib Hajimiragha; Leyla Sadighpour; Azam Sadat Mostafavi
Journal:  Dent Res J (Isfahan)       Date:  2020-12-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.