Implant failure in tibial nailing.

A L Ruiz1, W D Kealey, G F McCoy.   

Abstract

Between 1991 and 1995 five hundred and twenty tibial diaphyseal fractures were treated with intramedullary nailing in Northern Ireland. We retrospectively reviewed three hundred and thirty eight patients in three out of the four fracture units in Northern Ireland. Within this group there were three hundred and forty fractures and we identified twenty five episodes of implant failure. This group of patients had a higher percentage of open injuries with a higher degree of comminution and had been treated with smaller diameter nails when compared with the group of patients, who had no implant failure. Failure occurred most frequently at the transverse proximal locking screw when a single screw was used. Fractures in the distal third of the tibia had a higher incidence of nail breakage. We would recommend early elective dynamisation and the use of statically locked larger diameter nails, especially in comminuted or distal third fractures.

Entities:  

Mesh:

Year:  2000        PMID: 10775692     DOI: 10.1016/s0020-1383(00)00002-4

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

Review 1.  Timing issue in open fractures debridement: a review article.

Authors:  Alberto Jorge-Mora; Juan Rodriguez-Martin; Juan Pretell-Mazzini
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-04-04

2.  Intramedullary tibial nailing in distal third tibial fractures: distal locking screws and fracture non-union.

Authors:  Aso Mohammed; Ramaswamy Saravanan; Jason Zammit; Richard King
Journal:  Int Orthop       Date:  2007-04-05       Impact factor: 3.075

3.  A novel intramedullary nail for use in the treatment of supramalleolar malunion and nonunion: A preliminary report of three cases.

Authors:  Fatih Küçükdurmaz; Necdet Sağlam; Tuhan Kurtulmuş; Fuat Akpınar
Journal:  Acta Orthop Traumatol Turc       Date:  2016-11-11       Impact factor: 1.511

4.  Strength⁻Durability Correlation of Osteosynthesis Devices Made by 3D Layer Manufacturing.

Authors:  Yoshimitsu Okazaki; Emiko Gotoh; Jun Mori
Journal:  Materials (Basel)       Date:  2019-01-31       Impact factor: 3.623

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.