The effects of passive humidifier dead space on respiratory variables in paralyzed and spontaneously breathing patients.

R S Campbell1, K Davis, J A Johannigman, R D Branson.   

Abstract

BACKGROUND: Passive humidifiers have gained acceptance in the intensive care unit because of their low cost, simple operation, and elimination of condensate from the breathing circuit. However, the additional dead space of these devices may adversely affect respiratory function in certain patients. This study evaluates the effects of passive humidifier dead space on respiratory function.
METHODS: Two groups of patients were studied. The first group consisted of patients recovering from acute lung injury and breathing spontaneously on pressure support ventilation. The second group consisted of patients who were receiving controlled mechanical ventilation and were chemically paralyzed following operative procedures. All patients used 3 humidification devices in random order for one hour each. The devices were a heated humidifier (HH), a hygroscopic heat and moisture exchanger (HHME) with a dead space of 28 mL, and a heat and moisture exchanger (HME) with a dead space of 90 mL. During each measurement period the following were recorded: tidal volume, minute volume, respiratory frequency, oxygen consumption, carbon dioxide production, ratio of dead space volume to tidal volume (VD/VT), and blood gases. In the second group, intrinsic positive end-expiratory pressure was also measured.
RESULTS: Addition of either of the passive humidifiers was associated with increased VD/VT. In spontaneously breathing patients, VD/VT increased from 59 +/- 13 (HH) to 62 +/- 13 (HHME) to 68 +/- 11% (HME) (p < 0.05). In these patients, constant alveolar ventilation was maintained as a result of increased respiratory frequency, from 22.1 +/- 6.6 breaths/min (HH) to 24.5 +/- 6.9 breaths/min (HHME) to 27.7 +/- 7.4 breaths/min (HME) (p < 0.05), and increased minute volume, from 9.1 +/- 3.5 L/min (HH) to 9.9 +/- 3.6 L/min (HHME) to 11.7 +/- 4.2 L/min (HME) (p < 0.05). There were no changes in blood gases or carbon dioxide production. In the paralyzed patient group, VD/VT increased from 54 +/- 12% (HH) to 56 +/- 10% (HHME) to 59 +/- 11% (HME) (p < 0.05) and arterial partial pressure of carbon dioxide (PaCO2) increased from 43.2 +/- 8.5 mm Hg (HH) to 43.9 +/- 8.7 mm Hg (HHME) to 46.8 +/- 11 mm Hg (HME) (p < 0.05). There were no changes in respiratory frequency, tidal volume, minute volume, carbon dioxide production, or intrinsic positive end-expiratory pressure. DISCUSSION: These findings suggest that use of passive humidifiers with increased dead space is associated with increased VD/VT. In spontaneously breathing patients this is associated with an increase in respiratory rate and minute volume to maintain constant alveolar ventilation. In paralyzed patients this is associated with a small but statistically significant increase in PaCO2.
CONCLUSION: Clinicians should be aware that each type of passive humidifier has inherent dead space characteristics. Passive humidifiers with high dead space may negatively impact the respiratory function of spontaneously breathing patients or carbon dioxide retention in paralyzed patients. When choosing a passive humidifier, the device with the smallest dead space, but which meets the desired moisture output requirements, should be selected.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10771799

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  14 in total

1.  Home mechanical ventilation: a Canadian Thoracic Society clinical practice guideline.

Authors:  Douglas A McKim; Jeremy Road; Monica Avendano; Steve Abdool; Fabien Cote; Nigel Duguid; Janet Fraser; Fracois Maltais; Debra L Morrison; Colleen O'Connell; Basil J Petrof; Karen Rimmer; Robert Skomro
Journal:  Can Respir J       Date:  2011 Jul-Aug       Impact factor: 2.409

2.  Under-humidification and over-humidification during moderate induced hypothermia with usual devices.

Authors:  François Lellouche; Siham Qader; Solenne Taille; Aissam Lyazidi; Laurent Brochard
Journal:  Intensive Care Med       Date:  2006-05-23       Impact factor: 17.440

3.  Heat and moisture exchangers and heated humidifiers in acute lung injury/acute respiratory distress syndrome patients. Effects on respiratory mechanics and gas exchange.

Authors:  Indalecio Morán; Judith Bellapart; Alessandra Vari; Jordi Mancebo
Journal:  Intensive Care Med       Date:  2006-02-24       Impact factor: 17.440

4.  Influence of the humidification device during acute respiratory distress syndrome.

Authors:  Gwénaël Prat; Anne Renault; Jean-Marie Tonnelier; David Goetghebeur; Emmanuel Oger; Jean-Michel Boles; Erwan L'Her
Journal:  Intensive Care Med       Date:  2003-08-06       Impact factor: 17.440

Review 5.  Humidification and heating of inhaled gas in patients with artificial airway. A narrative review.

Authors:  Gustavo Adrián Plotnikow; Matias Accoce; Emiliano Navarro; Norberto Tiribelli
Journal:  Rev Bras Ter Intensiva       Date:  2018-03

6.  Low Tidal Volume Ventilation in Patients without Acute Respiratory Distress Syndrome: A Paradigm Shift in Mechanical Ventilation.

Authors:  Jed Lipes; Azadeh Bojmehrani; Francois Lellouche
Journal:  Crit Care Res Pract       Date:  2012-03-27

7.  The effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric anesthesia.

Authors:  Min A Kwon
Journal:  Korean J Anesthesiol       Date:  2012-05-24

Review 8.  Extracorporeal respiratory support in adult patients.

Authors:  Thiago Gomes Romano; Pedro Vitale Mendes; Marcelo Park; Eduardo Leite Vieira Costa
Journal:  J Bras Pneumol       Date:  2017 Jan-Feb       Impact factor: 2.624

Review 9.  Humidification during mechanical ventilation in the adult patient.

Authors:  Haitham S Al Ashry; Ariel M Modrykamien
Journal:  Biomed Res Int       Date:  2014-06-25       Impact factor: 3.411

10.  Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans.

Authors:  Jeanette Janaina Jaber Lucato; Thiago Marraccini Nogueira da Cunha; Aline Mela Dos Reis; Patricia Salerno de Almeida Picanço; Renata Cléia Claudino Barbosa; Joyce Liberali; Renato Fraga Righetti
Journal:  Rev Bras Ter Intensiva       Date:  2017 Apr-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.