Literature DB >> 10766367

Validation of a method for the determination of zolpidem in human plasma using LC with fluorescence detection.

P R Ring1, J M Bostick.   

Abstract

A sensitive and selective high-performance liquid chromatography (HPLC) method was developed for the determination of zolpidem in human plasma. Zolpidem and the internal standard (trazodone) were extracted from human plasma that had been made basic. The basic sample was loaded onto a conditioned Bond Elut C18 cartridge, rinsed with water and eluted with methanol. Forty microliters were then injected onto the LC system. Separation was achieved on a C18 column (150 x 4.6 mm, 5 microm) with a mobile phase composed of acetonitrile:50 mM potassium phosphate monobasic at pH 6.0 (4:6, v/v). Detection was by fluorescence, with excitation at 254 nm and emission at 400 nm. The retention times of zolpidem and internal standard were approximately 4.7 and 5.3 min, respectively. The LC run time was 8 min. The assay was linear in concentration range 1-400 ng/ml for zolpidem in human plasma. The analysis of quality control samples for zolpidem (3, 30, and 300 ng/ml) demonstrated excellent precision with relative standard deviations (RSD) of 3.7, 4.6, and 3.0%, respectively (n = 18). The method was accurate with all intraday (n = 6) and overall (n = 18) mean concentrations within 5.8% from nominal at all quality control sample concentrations. This method was also performed using a Gilson Aspec XL automated sample processor and autoinjector. The samples were manually fortified with internal standard and made basic. The aspec then performed the solid phase extraction and made injections of the samples onto the LC system. Using the automated procedure for analysis, quality control samples for zolpidem (3, 30, and 300 ng/ml) demonstrated acceptable precision with RSD values of 9.0, 4.9, and 5.1%, respectively (n = 12). The method was accurate with all intracurve (n = 4) and overall (n = 12) mean values being less than 10.8% from nominal at all quality control sample concentrations.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10766367     DOI: 10.1016/s0731-7085(99)00311-8

Source DB:  PubMed          Journal:  J Pharm Biomed Anal        ISSN: 0731-7085            Impact factor:   3.935


  4 in total

1.  Relationship between zolpidem concentrations and sleep parameters in pediatric burn patients.

Authors:  Chris Stockmann; Michele M Gottschlich; Daniel Healy; Jane C Khoury; Theresa Mayes; Catherine M T Sherwin; Michael G Spigarelli; Richard J Kagan
Journal:  J Burn Care Res       Date:  2015 Jan-Feb       Impact factor: 1.845

2.  Spectrophotometric methods for determination of zolpidem tartrate in tablet formulation.

Authors:  Rajiv Chomwal; Amit Kumar; Anju Goyal
Journal:  J Pharm Bioallied Sci       Date:  2010-10

3.  Zolpidem : Forensic aspects for the toxicologist and pathologist.

Authors:  Timothy P Rohrig; Christine M Moore
Journal:  Forensic Sci Med Pathol       Date:  2005-06       Impact factor: 2.456

4.  Development and Validation of a Stability-Indicating Capillary Electrophoresis Method for the Determination of Zolpidem Tartrate in Tablet Dosage Form with Positive Confirmation using 2D- and 3D-DAD Fingerprints.

Authors:  Khaldun M Al Azzam; Lee Kam Yit; Bahruddin Saad; Hassan Shaibah
Journal:  Sci Pharm       Date:  2014-02-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.