Literature DB >> 10760112

Influence of convection on small molecule clearances in online hemodiafiltration.

A Ficheux1, A Argilés, H Mion, C M Mion.   

Abstract

BACKGROUND: Dialysis efficacy is mostly influenced by dialyzer clearance. Urea clearance may be estimated in vitro by total ion clearance, which can be obtained by conductivity measurements. We have previously used this approach to assess in vitro clearances in a system mimicking predilutional and postdilutional online hemodiafiltration with a wide range of QD, QB, and ultrafiltration rates. Our current study elaborates on a formula that allows the prediction of the influence of ultrafiltration on small molecule clearances, and validates the mathematical approach both experimentally in vitro and clinically in vivo data.
METHODS: Two conductivimeters in the dialysate side of an E-2008 Fresenius machine were used. HF80 and HF40 polysulfone dialyzers were used; reverse osmosis water and dialysate were used for blood and dialysate compartments, respectively. Study conditions included QB of 300 and 400 mL/min and QD of 500 and 590 mL/min, with a range of ultrafiltration rate from 0 to 400 mL/min in postdilutional hemodiafiltration and to 590 mL/min in predilutional hemodiafiltration. Urea clearances were determined in the in vivo studies, which included 0, 50, 100, and 150 mL/min ultrafiltration rates.
RESULTS: The ultrafiltration rate and clearance were significantly correlated (R > 0.9, P < 0.001) and fitted a linear model (P < 0.001) in all of the experimental conditions. The following formula fitted the experimental points with an error <2% for both postdilutional and predilutional online diafiltration in vitro, respectively. K = K0 + [(QB - K0)/(QB)] x ultrafiltration rateK = K0 + [((QD x QB)/(QB + QD) - K0)/QD] x ultrafiltration rate where K is the clearance; K0 is the clearance with nil ultrafiltration rate; QD is the total dialysate produced (in commercial HDF, QD = QDi + Qinf). Since weight loss was maintained at 0, ultrafiltration rate = infusion flow. QB is the "blood" line flow. The formula was also verified in vivo in clinical postdilutional hemodiafiltration with a QB taking into account the cellular and water compartments. DISCUSSION: In vitro, by simply determining the clearance in conventional dialysis, the total clearance for any ultrafiltration rate may be estimated in both predilutional and postdilutional online diafiltration with an error of less than 2%. The same applies to in vivo postdilutional hemodiafiltration when the formula takes into account the cellular and water composition of blood.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10760112     DOI: 10.1038/sj.ki.4495463

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  2 in total

1.  Consequences of increasing convection onto patient care and protein removal in hemodialysis.

Authors:  Nathalie Gayrard; Alain Ficheux; Flore Duranton; Caroline Guzman; Ilan Szwarc; Fernando Vetromile; Chantal Cazevieille; Philippe Brunet; Marie-Françoise Servel; Àngel Argilés; Moglie Le Quintrec
Journal:  PLoS One       Date:  2017-02-06       Impact factor: 3.240

2.  A reliable method to assess the water permeability of a dialysis system: the global ultrafiltration coefficient.

Authors:  A Ficheux; N Gayrard; F Duranton; C Guzman; I Szwarc; F Vetromile; P Brunet; M F Servel; A Argilés
Journal:  Nephrol Dial Transplant       Date:  2017-02-01       Impact factor: 5.992