Literature DB >> 10755379

Measurement of both protein-bound and total S-2-(3-aminopropylamino)ethanethiol (WR-1065) in blood by high-performance liquid chromatography.

H S Bonner1, L M Shaw.   

Abstract

A high-performance liquid chromatographic method for automated analysis of both protein-bound and total S-2-(3-aminopropylamino)ethanethiol (WR-1065) in blood has been developed in our laboratory. WR-1065 is the active thiol metabolite of the radio- and chemo-protector drug amifostine (WR-2721). Using WR-1065 quality control levels over the experimental range: 7.0, 45.0 and 85.0 micromol/l spiked into plasma, method validation for total WR-1065 included between-run assessment of imprecision (SD/C.V.%: 1.11/16.7%, 6.58/15.5% and 9.24/11.3%, respectively) and % accuracy (94.7, 106.0 and 97.2%).

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10755379     DOI: 10.1016/s0378-4347(99)00523-x

Source DB:  PubMed          Journal:  J Chromatogr B Biomed Sci Appl        ISSN: 1387-2273


  1 in total

1.  Development and Validation of an HPLC Method for Determination of Amifostine and/or Its Metabolite (WR-1065) In Human Plasma Using OPA Derivatization and UV Detection.

Authors:  Nasim Samiei; Seyed Mohsen Foroutan; Alireza Shafaati; Afshin Zarghi
Journal:  Iran J Pharm Res       Date:  2015       Impact factor: 1.696