Tesio catheter: radiologically guided placement, mechanical performance, and adequacy of delivered dialysis.

S Perini1, J M LaBerge, J M Pearl, H L Santiestiban, H E Ives, R S Omachi, M Graber, M W Wilson, S R Marder, B R Don, R K Kerlan, R L Gordon.   

Abstract

PURPOSE: Tunneled catheters are an alternative means of vascular access for patients in need of hemodialysis who cannot undergo dialysis through a surgical shunt. This study was undertaken to evaluate the performance of the Tesio dialysis catheter.
MATERIALS AND METHODS: A prospective study of the Tesio catheter was performed. Follow-up data regarding catheter function and adequacy of dialysis were obtained from nine hemodialysis facilities.
RESULTS: Seventy-nine Tesio catheters were placed in 71 patients. Immediate technical success was 99% (78 of 79 catheters). The procedure complication rate was 9% (seven catheters). Only two complications required intervention: one fatal air embolism and one chest wall hematoma. Sixty-seven catheters in 60 patients were followed up for a total of 4,367 catheter days. Overall, catheter-related infection occurred in 9% (six of 67 catheters). Primary catheter patency was 87% at 1 week, 82% at 1 month, 72% at 3 months, and 66% at 6 months. Mean blood flow was 286 mL/min immediately after insertion, 301 mL/min at 3 months, and 306 mL/min at 6 months. Adequate dialysis dose as reflected by a urea reduction ratio of 60 or more or a urea kinetic modeling, or Kt/V, value of 1.2 or more was observed on at least one occasion for 74% and 76% of catheters, respectively.
CONCLUSION: The Tesio catheter is a reasonable means of vascular access for patients who undergo dialysis but are not candidates for surgical shunt placement.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10751478     DOI: 10.1148/radiology.215.1.r00mr43129

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105