Abstract
The sensitivities (in m) of bare LR115 detectors and detectors in diffusion chambers to 222Rn and 220Rn chains are calculated by the Monte Carlo method. The partial sensitivities of bare detectors to the 222Rn chain are larger than those to the 220Rn chain, which is due to the higher energies of alpha particles in the 220Rn chain and the upper energy limit for detection for the LR115 detector. However, the total sensitivities are approximately equal because 220Rn is always in equilibrium with its first progeny, which is not the case for the 222Rn chain. The total sensitivity of bare LR115 detectors to 222Rn chain depends linearly on the equilibrium factor. The overestimation in 222Rn measurements with bare detectors caused by 220Rn in air can reach 10% in normal environmental conditions. An analytical relationship between the equilibrium factor and the ratio between track densities on the bare detector and the detector enclosed in chamber is given in the last part of the paper. This ratio is also affected by 220Rn, which can disturb the determination of the equilibrium factor.Entities: