Overestimation of hemoglobin in a patient with an IgA-kappa monoclonal gammopathy.

W L Roberts1, J D Fontenot, C M Lehman.   

Abstract

A patient with multiple myeloma had an automated blood count performed on a Coulter STK-S counter that repeatedly failed internal limits for both mean corpuscular hemoglobin and mean corpuscular hemoglobin concentration. The calculated hematocrit agreed with a spun hematocrit, suggesting that the hemoglobin concentration was being overestimated by the automated counter. Measurement of the plasma hemoglobin concentration of the sample, which showed no visible hemolysis, gave a hemoglobin concentration of 32 g/L on the STK-S analyzer. Correction of the whole blood hemoglobin using the plasma hemoglobin gave a value consistent with the hematocrit. The corrected mean corpuscular hemoglobin and mean corpuscular hemoglobin concentration values were within standard limits. This patient's paraprotein was characterized as IgA-kappa and was present at a concentration of 61 g/L. The hemoglobin concentration measured on whole blood by Sysmex NE 8000 and Technicon H*1E autoanalyzers agreed reasonably well with the corrected result from the STK-S.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10747323     DOI: 10.5858/2000-124-0616-OOHIAP

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  2 in total

1.  Evaluation of Double M-Band on Serum Protein Electrophoresis Simulating Biclonal Gammopathy: A Case Report.

Authors:  Pallavi Jain; Reeta Choudhary; Arun Kumar Harith; Charu Yadav
Journal:  Indian J Clin Biochem       Date:  2021-04-08

Review 2.  Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution.

Authors:  Gene Gulati; Guldeep Uppal; Jerald Gong
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.