Literature DB >> 10744305

Devices for ischemic preconditioning of the pedicled groin flap.

M H Cheng1, H C Chen, F C Wei, S W Su, S H Lian, E Brey.   

Abstract

BACKGROUND: Early division of the pedicled groin flap can be achieved by using ischemic preconditioning. The goal of this study was to investigate the devices available for ischemic preconditioning and determine which device is the most effective and results in lowest patient discomfort.
METHODS: Rubber bands, custom-made Orthoplast sheets, long-nose locking pliers, intestinal clamps, and a pneumatic tourniquet device were used for ischemic preconditioning on 13 patients who sustained severe hand injuries with reconstruction of pedicled groin flaps. The devices were compared by using laser Doppler flowmetry and the patient's local pain levels.
RESULTS: Twelve of 13 flaps were successfully divided at a mean period of 8.3 days by using a custom-made Orthoplast sheet or a pneumatic tourniquet device. All devices except the rubber bands could result in a biologic zero flow level.
CONCLUSION: The pneumatic tourniquet device is the most desirable ischemic preconditioning device, having the advantages of excellent ischemic effect, easy application, and minimal discomfort.

Entities:  

Mesh:

Year:  2000        PMID: 10744305     DOI: 10.1097/00005373-200003000-00035

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  1 in total

1.  A modified bowel clamp technique for ischemic preconditioning.

Authors:  Jennifer Fong Ha; Paul Wilson
Journal:  Ochsner J       Date:  2009
  1 in total