Accuracy of coronary flow measurements performed by means of Doppler wires.

P Doriot1, P Dorsaz, L Dorsaz, P Chatelain.   

Abstract

To estimate in patients the accuracy of coronary flow measurements performed by means of 0.014" Doppler wires, the time-averaged maximal blood velocity (APV) was recorded in the 3 branches of 36 angiographically normal coronary artery bifurcations selected in 21 patients undergoing cardiac catheterization for various diseases. Contrast medium injections filmed under two incidences allowed identification of the 3 sample volume locations and computing of the 3 corresponding vessel cross-sectional areas (CSA) at subsequent data analysis. Multiplication of the velocities APV/2 (range: 3 to 20.5 cm/s) by the CSA (obtained by averaging the two calibrated vessel diameters; range: 1.6 to 5.4 mm) yielded 108 flow rates (range: 5.4 to 169 mL/min). The average relative flow error was then estimated using the continuity equation (Q(in) = Q(out,1) + Q(out, 2)) and the central limit theorem. The result was that the relative flow error decreased from 30% at Q = 30 mL/min to 13% at Q = 160 mL/min. We conclude that coronary flow measurements are reasonably accurate, except perhaps for very low flows.

Entities:  

Mesh:

Year:  2000        PMID: 10722911     DOI: 10.1016/s0301-5629(99)00133-7

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  3 in total

1.  Post-stenotic coronary blood flow at rest is not altered by therapeutic doses of the oral antidiabetic drug glibenclamide in patients with coronary artery disease.

Authors:  T Reffelmann; H G Klues; P Hanrath; E R Schwarz
Journal:  Heart       Date:  2002-01       Impact factor: 5.994

2.  Pulsatile arterial wall-blood flow interaction with wall pre-stress computed using an inverse algorithm.

Authors:  Ashish Das; Anup Paul; Michael D Taylor; Rupak K Banerjee
Journal:  Biomed Eng Online       Date:  2015-01-09       Impact factor: 2.819

3.  Compound ex vivo and in silico method for hemodynamic analysis of stented arteries.

Authors:  Farhad Rikhtegar; Fernando Pacheco; Christophe Wyss; Kathryn S Stok; Heng Ge; Ryan J Choo; Aldo Ferrari; Dimos Poulikakos; Ralph Müller; Vartan Kurtcuoglu
Journal:  PLoS One       Date:  2013-03-13       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.