| Literature DB >> 10720237 |
M Kele1, G Guiochon.
Abstract
The reproducibility of retention data and band profile characteristics was investigated for a series of columns packed with Luna C18 (2), a silica-based, reversed-phase adsorbent. High precision data were obtained and statistically compared among five columns from the same batch (column-to-column reproducibility) and nine columns from as many different batches (batch-to-batch reproducibility). These data were acquired under five different sets of chromatographic conditions, for a group of 30 neutral, acidic, and basic compounds selected as probes following an experimental protocol previously described. This work is part of a study on the precision of chromatographic analyses. Its purpose is to illustrate the contribution of the stationary phase reproducibility to this precision.Entities:
Mesh:
Year: 2000 PMID: 10720237 DOI: 10.1016/s0021-9673(99)01092-4
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759