A comparison of modified Howland circuits as current generators with current mirror type circuits.

P Bertemes-Filho1, B H Brown, A J Wilson.   

Abstract

Multi-frequency electrical impedance tomography (EIT) systems require stable voltage controlled current generators that will work over a wide frequency range and with a large variation in load impedance. In this paper we compare the performance of two commonly used designs: the first is a modified Howland circuit whilst the second is based on a current mirror. The output current and the output impedance of both circuits were determined through PSPICE simulation and through measurement. Both circuits were stable over the frequency ranges 1 kHz to 1 MHz. The maximum variation of output current with frequency for the modified Howland circuit was 2.0% and for the circuit based on a current mirror 1.6%. The output impedance for both circuits was greater than 100 kohms for frequencies up to 100 kHz. However, neither circuit achieved this output impedance at 1 MHz. Comparing the results from the two circuits suggests that there is little to choose between them in terms of a practical implementation.

Mesh:

Year:  2000        PMID: 10719993     DOI: 10.1088/0967-3334/21/1/301

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  4 in total

1.  Design of Bioimpedance Spectroscopy Instrument With Compensation Techniques for Soft Tissue Characterization.

Authors:  Robert E Dodde; Grant H Kruger; Albert J Shih
Journal:  J Med Device       Date:  2015-06       Impact factor: 0.582

2.  An Implantable Wireless Neural Interface System for Simultaneous Recording and Stimulation of Peripheral Nerve with a Single Cuff Electrode.

Authors:  Ahnsei Shon; Jun-Uk Chu; Jiuk Jung; Hyungmin Kim; Inchan Youn
Journal:  Sensors (Basel)       Date:  2017-12-21       Impact factor: 3.576

3.  A Low-Power Stable Wideband Current Source for Acupuncture Point Skin Impedance Measurements.

Authors:  Changpei Qiu; Tianxia Zhao; Qiuping Li; Xin'an Wang; Kanglin Xiao; Bo Wang
Journal:  J Healthc Eng       Date:  2021-01-06       Impact factor: 2.682

4.  Frequency-division multiplexing for electrical impedance tomography in biomedical applications.

Authors:  Yair Granot; Antoni Ivorra; Boris Rubinsky
Journal:  Int J Biomed Imaging       Date:  2007
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.